background image

IM12B06J01-40E-E

2.6 Sensor Dimensions

Dimension according to DIN 19263:2007-05

Criteria 

: 120 mm version

L (below plug head) : 120 ± 2 mm

Ø shaft 

: 11.9 ± 0.1 mm 

Criteria 

: 225 mm version

L (below plug head) : 225 ± 2 mm

Ø shaft 

: 11.9 ± 0.1 mm 

Concentricity 

:  < 0,25 mm (centerline plug 

head - 25 mm from tip)

Perpendicularity 

:  < 0,5 mm (plug head  

- 25 mm from tip)

Cilindricity 

:  < 0,5 mm (centerline 

sensor)

Criteria 

: All types

All measuring elements within 25mm from tip.

Figure 1

2.7  Equipment and protective systems  

for use in potentially explosive 

atmospheres 

ATEX: 

  II 1 G Ex ia IIC T3...T6 Ga

 

 EU-type Examination Certificate 

Number: DEKRA 11ATEX0014 X

IECEx:

  Ex ia IIC T3...T6 Ga

 

 IECEx Certificate of Conformity 

Cert. No: IECEx DEK 11.0064X

pH sensor series SC25V for connection to a 

certified associated pH transmitter are used 

for measurement of pH and temperature.

Ambient temperature range:

-40°C to +40°C for temp. class T6,

-40°C to +55°C for temp. class T4 and T5,

-40°C to +105°C for temp. class T3,

Electrical data

For the models without ID-chip the I/O 

signals are from/to an associated intrinsically 

safe certified pH transmitter (e.g. Yokogawa 

transmitter Model FLXA21 series or 

Yokogawa transmitter Model PH202S series).

For the models with ID-chip the I/O signals 

are from/to an associated intrinsically safe 

certified pH transmitter, Yokogawa Smart 

Adapter Model SA11-P1.

Sensor input circuits:

In type of protection intrinsic safety Ex ia IIC, 

only for connection to a certified intrinsically 

safe circuit, with maximum values:  

Ui=18 V; Ii=170 mA; Pi=0.4 W; Li=0 mH; 

Ci=0 nF (without ID-chip) 

Ci=0.4 nF (with ID-chip)

Special conditions for safe use:

For ambient temperature see above.

Electrostatic charges of the sensor enclosure 

parts and label shall be avoided, especially 

in the case that the process medium is non-

conductive. Use a damp cloth for cleaning 

the equipment.

Note:

  When the sensor has been connected 

to none intrinsically safe equipment 

or equipment which exceeds the 

restrictions regarding the sensor input 

circuit, the sensor is not suitable for 

intrinsically safe use anymore.

120

±2

.0

  /

  22

2.

0

PG13.5

DIN: 40430

Wrench 19

Variopin

ø

11.9±0.1

ø

10.8

<2

5.

0

S
N

C

3

2
P

5

6

V
0

-A
00

IP67

173

±2

.0

 / 

27

2.

0

Dimensions in mm

Summary of Contents for SC25V-AGP25-120

Page 1: ...IM12B06J01 40E E 4th edition Instruction Manual YOKOGAWA Model SC25V Combined 12mm sensor pH Ref LE and Temperature...

Page 2: ...2 IM12B06J01 40E E...

Page 3: ...2 5 Dynamic specifications 5 2 6 Sensor Dimensions 6 2 7 Equipment and protective systems for use in potentially explosive atmospheres 6 3 INSTALLATION 7 3 1 Typical installation 7 3 2 Preparing the...

Page 4: ...typical warranty period and the conditions of sales relating to the original purchase order should be consulted Damage caused by wear and tear inadequate maintenance corrosion or by the effects of ch...

Page 5: ...emperature control 2 4 Operating range pH 0 to 14 Temperature Type G glass 10 to 80 C 14 to 176 F Type L glass 15 to 130 C 59 to 266 F Pressure 0 to 10 bar 0 to 145 psi Conductivity 10 S cm Storage te...

Page 6: ...hout ID chip the I O signals are from to an associated intrinsically safe certified pH transmitter e g Yokogawa transmitter Model FLXA21 series or Yokogawa transmitter Model PH202S series For the mode...

Page 7: ...he sensor upright for 24 hours Before mounting the sensor it should be validated with buffer solutions and calibrated when necessary The specific calibration procedure is described in the Instruction...

Page 8: ...1523JB JD PG13 5 adapter K1523JA JC Conductivity Flow Fitting FF40 SC4A option FF K1523JA K1598AC Flow fitting FF40 with K1523JA JC Adapter to fit sensors with a PG13 5 process connection in FF40 FS40...

Page 9: ...le etc 5 Rinse sensor thoroughly with clean water Carefully slap out residual fluid drops 6 Immerse the sensor in the second buffer pH 4 01 buffer or pH 9 18 buffer recommended 7 Repeat Step 3 Most in...

Page 10: ...O RINGS EPDM 11X3 6 PCS K1500BZ O RINGS VITON 11X3 6 PCS K1500GR O RINGS SILICON 11X3 8PCS K1524AA Set of O ring 11x3 and slide ring Ryton FP20 R12 Blind plug set for 1 hole 7 2 Adapters Prod No Desc...

Page 11: ...o o o o o o o x Ammonia in water 30 x x x o o o x x o o o o o o o o x Ammonium chloride sat o o o o o o o o o o o o o o o o o o Zinc chloride 50 o o o o o o o o o o o o o o o o o o Iron III chloride 5...

Page 12: ...herlands www yokogawa com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newna...

Reviews: