< 2. Installation and Wiring >
2-2
IM 19C1B3-01E
2.1.2 Installing Procedure
The ultrasonic oscillator may be mounted on panel, a wall surface or a pipe (2-inch, OD: 60.5 mm).
Hardware corresponding to the type of mounting used is required to install the PUS400G. Mounting
hardware is attached only when requested.
In the case of PUS400G-NN-KC (for Korea), a noise filter assembly is attached. Please refer to a
figure of external dimensions for the attachment means of the noise filter assembly.
(1)
Panel Mounting
Figure 2.2 illustrates the installing procedure for the ultrasonic oscillator using panel mounting bracket.
Panel
Panel mounting bracket
Panel retaining screw
Figure 2.2
Mounting Method for Ultrasonic Oscillator Using Panel Mounting Bracket
To mount the PUS400G on a panel, cut out the panel in accordance with Figure 2.3.
173
+1.1
0
156
+1.1
0
121
121
138
138
Unit : mm
Figure 2.3
Panel Cutout Dimensions