IM 12B7J1-01E
< 2. Installation >
2-8
F2.9.ai
Protector
O-ring
pH Sensor
Figure 2.9
Installing Sensor in Submersion Holder
(7) Close the waterproof cap, and attach the holder to the arm pipe, flange or cleaner holder
completely.
2.2.3
Installing Sensor in Flow-through Holder
To install the sensor in a flow-through holder, proceed as follows:
(1) Connect the sensor cable to the associated instrument. First, remove the sensor fixing nut
and pass the sensor cable through the nut.
pH Sensor
Sensor Fixing Nut
Figure 2.10
Preparation for Sensor Cable Connection
Properly connect the sensor cable by referring to Section 2.3.
(2) Mount the liquid junction on the sensor body. Peel off the seal covering the liquid junction
mounting hole in the sensor body. Screw the liquid junction softly into the hole by turning it
for two or three turns.
12th Edition : Dec.14,2018-00