5-8 Parameter setting
IM 12B6B3-E-H
5-2. Commissioning mode
In order to obtain peak performance from the EXA, you must set it up for each custom application.
Setpoints
Alarms are set by default
S1 - high process alarm
S2 - low process alarm
S3 - WASH
S4 - FAIL
The setpoints are at arbitrary default value. Therefore, you must set these to mean-
ingful values, or set them to off. (See service codes 40 to 49 and user interface
codes 50 to 59).
Output ranges
mA output 1 is set as default to 0 - 14 pH
mA output 2 is set as default to 0 - 100 °C
For enhanced resolution in more stable measuring processes, it may be desirable
to select 5 - 10 pH range, for example, and maybe 0 - 25 °C temperature range.
Service codes 30 to 39 can be used to choose other output parameters on mA
output 2. Choose from ORP, temperature, rH or PI control.
Hold
The EXA converter has the ability to “hold” the output during maintenance periods.
This parameter should be set up to hold the last measured value, or a fixed value
to suit the process.
Wash cleaner
The EXA can be set up to control a wash cleaner. When using this function, the
timings must be configured for interval, wash and recovery periods.
Service
This selection provides access to the service menu.
What follows are pictorial descriptions of typical frontplate pushbutton sequences for each parameter set-
ting function. By following the simple YES/NO prompts and arrow keys, users can navigate through the
process of setting range, setpoints, hold, wash and service functions.