Input voltage range
35V
–35V
35V
–35V
Use it within this range.
Vin+
Positive signal input terminal
Vin-
Negative signal input terminal
General Specifications
Supply voltage range Interface
Standard supply voltage ± 5V, within ±5%
(Power is supplied to the probe through a
dedicated terminal. Connect the probe to a digital
oscilloscope with a compatible terminal.)
Storage altitude
3000 m or less
Operating altitude
2000 m or less
Standard operating environment
Temperature range 23 ± 5°C
Humidity range
6
55 ± 10%RH
Operating environment
Temperature range 5 to 40°C
Humidity range
6
20 to 80%RH
Storage environment
Temperature range –20 to 60°C
Humidity range
6
20 to 80%RH
Calibration period
1 year
Warm-up time
At least 30 minutes
Total length
Approx. 1.2 m
Weight
Approx. 90 g
6 No condensation.
Compliant Standards
EMC standards
Emission
EN61326-1 Class A
EN55011 Class A, Group1
EMC standards of Australia and New Zealand
EN55011 Class A, Group1
Immunity
EN61326-1 Table 2 (for use in industrial locations)
Influence in immunity test environment
|Noise increase| ≤ 2 V
7
Environmental standards
EU RoHS Directive compliant
7 Test conditions
Frequency bandwidth limit 20 MHz, using a DL9000 series digital oscilloscope
with the input impedance set to 50 Ω, and both plus and minus probe tip inputs
connected (terminated) to 50 Ω.