Product Specifications
Electrical Specifications
(The electrical specifications are based on standard operating environment after
30-minute warm-up.)
Frequency bandwidth
DC to 1 GHz (–3 dB or higher)
Attenuation ratio and DC
voltage accuracy
1
50:1, within ±2% of the differential input voltage
2
(into 50-Ω load, excluding oscilloscope errors. See
appendix 2.)
Input capacitance
Approx. 1.1 pF (relative to ground, typical value
5
)
Input resistance
Within 1 MΩ ± 3% (relative to ground)
Output impedance
Approx. 50 Ω (typical value
5
)
Maximum operating input
voltage range
±35 V (DC + ACpeak)
Maximum differential input
voltage range
±25 V (DC + ACpeak)
Maximum non-destructive
input voltage
3
±100 V (instantaneous)
Maximum non-destructive
continuous input voltage
±35 V (DC + ACpeak)
Rise time
350 ps or less (excluding characteristics of the
oscilloscope, typical value
5
)
Residual noise
500 μVrms or less (at the probe output, typical value
5
)
Residual offset
4
Within ±10 mV (after adjustment)
Common mode rejection
ratio
DC to 1 MHz:
–35 dB or less
1 MHz to 10 MHz
–30 dB or less
10 MHz to 100 MHz: –26 dB or less
100 MHz to 300 MHz: –20 dB or less
1 Excludes residual offset voltage.
2 Under standard operating environment (23°C ± 5°C).
3 Maximum instantaneous voltage that will not break the probe. It is not the rated
voltage that you can use continuously. For continuous usage, use it within the
maximum operating input voltage range.
4 When 0 V is applied to both positive and negative input terminals.
5 Typical values represent typical or average values. They are not strictly warranted.
1
10
100
1
10
100
1000
Frequency (MHz)
Derating of the input voltage by frequency
Maximum non-destructive
continuous input voltage range
(V peak)