6.1.2
Setting Secondary Output “oUt2”
Operation: Hold down the [MODE] key for 2 seconds, select “oUt2” with the [▲], [▼] keys,
and press the [ENT] key.
With respect to the secondary output, select one of the three types of voltage output: 0-1 V is “1,”
0-5 V is “5,” and 0-10 V DC is “10.” Figure 6.3 shows an example of setting 0-1 V DC “1.”
CAL
rnG
oUt2
HoLd
▼↓↑▲
▼↓↑▲
▼↓↑▲
▼↓↑▲
[ENT]
1
oUt2
PPM
5
oUt2
10
oUt2
▼↓↑▲
▼↓↑▲
▼↓↑▲
▼↓↑▲
[ENT]
1
oUt2
PPM
[ENT]
MODE
(0-10 V D C )
(0-5 V D C )
%
%
To “rEV”
Hold down
for 2 seconds.
To “10”
0-1 V DC is confirmed.
Return to “oUt2.”
To “1”
Figure 6.3
Setting Secondary Output “oUt2”
6.1.3
Setting HOLD Function “HoLd”
Operation: Press the [MODE] key for 2 seconds, select HoLd with the [▲], [▼] keys, and
press the [ENT] key.
Set the output state in the maintenance mode. Select either “non” or “PrEV.” The former option
means no hold, and the latter means hold the previous value. The following is an example of no
hold.
HoLd
▼↓↑▲
CAL
▼↓↑▲
▼↓↑▲
[ENT]
HoLd
PPM
HoLd
▼↓↑▲
▼↓↑▲
▼↓↑▲
[ENT]
HoLd
PPM
[ENT]
MODE
non
PrEV
non
To “rEV”
Hold down
for 2 seconds.
To “PrEV”
No hold is confirmed.
Return to “HoLd.”
(Hold the previous value.)
To “non”
%
%
Figure 6.4
Setting HOLD “HoLd”