Item

Limit

Test condition

Response time

3 sec or less

From the instant the resistors whose values 
correspond to central indication and  
zero indications are abruptly connected,  
to when the pointer reaches a level within tolerance

Friction

No friction is recognized

---

Effects of 
inclination

2% or less of the scale length

Deviation from the infinity indication when the tester 

is tilted from a horizontal position by 30 degrees

Effect of 
temperature

(1) 5% or less

(2) 0.7% or less of scale length

(1) Deviation from central indication when ambient 

temperature is varied from 20°C by ±20°C

(2) Deviation from infinity and zero indications when 

ambient temperature is varied from 20°C by ±20°C

Effect of humidity

Within tolerance

When the tester is left for 1 hour  

with the humidity at 90% RH

Effect of external  

magnetic field

3% or less of indication

The change of indication when an external field of 

400 A/m DC is applied in the most affected direction

Effect of  

AC component of 

 

measuring 

terminal voltage

10% or less of indication

A change when a capacitor with 5μF ±10% is 

connected in parallel to a resistor corresponding to 
central indication connected

Insulation 
resistance

50MΩ or more 

 

(between electric circuits and  
outer case)

When tested at 500 V if the rated measuring voltage 

is 500 V or less or at a voltage identical to 

 

the rated voltage if the rated voltage exceeds 500 V

Dielectric  

withstand voltage

There must not be an abnormality. 
(between electric circuits and  
outer case)

When a sine wave, or the like, is applied between 
the electric circuits and the outer case at  

5550 VAC and 50/60 Hz for 1 minute

Resistance to 

vibration

The specifications for tolerances, 

friction, and the effect of tilting  
must be met. 
There are no mechanical and 
electrical damage.

When a vibration frequency of 16.7 Hz and 

 

peak-to-peak amplitude of 4 mm is applied in 

 

the direction of the axis of the moving part for 1 hour

Resistance to  

shock

The specifications for tolerances, 

friction, and the effect of tilting
must be met. 
There are no mechanical and 
electrical damage.

When shock of 1000 m/s

2

 is applied twice in  

the direction of axis of the moving part and in  
a direction perpendicular to the axis

Endurance

The specifications for tolerances 

and friction must be met.

When a resistor corresponding to the central 
indication is connected across the measuring 
terminals and the power switch is repeatedly  
turned on and off 10 000 cycles at a rate of about 

300 cycles/hour

Input over-voltage

 

protection

There must not be an abnormality.

When an AC voltage of 50/60 Hz at 1.2 times 

 

the rated measuring voltage is applied to  
the measuring terminals for 10 seconds

General Specifications

Operation temperature and humidity:

0 to 40°C at 90% RH or less (no condensation)

Storage temperature and humidity:

-10 to 60°C at 70% RH or less (no condensation)

Battery:

AA-size (4), [R6]

External dimensions:

Approx. 125 (W) × 103 (H) × 52.5 (D) mm

Weight:

Approx. 400 g (main unit and batteries only)

Approx. 600 g
(main unit, batteries, protective cover, earth probe and line probe)

Safety standards:

EN 61010-1, EN 61010-2-030, EN 61010-031

Measurement category III (CAT III) 600 V

Insulation class II, 

Indoor use, altitude 2000 m or less, pollution degree 2

EMC Standard:

EN 61326-1 Class B, EN 61326-2-2

EMC Regulatory Arrangement in Australia and New Zealand

EN55011 Class B  Group 1

Korea Electromagnetic Conformity Standard

( 한국 전자파적합성기준 )

Effect of radiation immunity

(at the strength of radio-frequency 

electromagnetic field of 3 V/m)

Insulation resistance measurement: 

 

1st effective measuring range:  ±10% of rdg

 

2nd effective measuring range: ±20% of rdg

Infinity and zero indications: ±1.4% or less of scale length

AC voltage measurement: ±10% of F.S.

Environmental standard:

EN 50581

Monitoring and control instruments including industrial monitoring 

and control instruments

■ Standard Accessories

Name

Model No.

Quantity

Protection cover

93013

1

Shoulder strap

99005

1

Line probe

98001

1

Earth probe

98002

1

Batteries

---

4

User’s manual

IM MY10-E

1

■ Optional Accessories

Name

Model No.

Description

1. Spare probe tip for the line probe

(Model 98001)

99011

105 mm, breaker pin

2. Hard case

93015

Houses both the main unit,  
the line probe and the earth probe.

3. Accessory bag

B9108XA

Soft case, 

approx. 100 (W)  190 (H) 40 (D) mm

 

■ External Dimensions

125 (4.92)

103 (4.06)

52.5 (2.07)

Main unit

Line probe (98001)

Earth probe 

98002

 

Cap

Unit: mm

(approx. inch)

Cable length: approx. 1200 mm (approx. 47.24 inch)

Note

 

If the breaker pin (99011) is attached on the Line probe, detach the cap from Line probe.

10. Maintenance

■ Storage Conditions

• Temperature and humidity:-10°C to 60°C at 70% RH or less

• Remove the batteries before storing the tester.

• Avoid storing the tester in a location where there is:

  moisture; exposure to direct sunlight;
  a high-temperature heat source nearby;
  exposure to severe mechanical vibrations;
  a large amount of dust and/or salt, or a corrosive gas.

■ Removal of Dirt

Do not use volatile solvents (such as paint thinners or benzine) as they are likely to cause discoloration. 
Wipe off dirt with a cloth dampened water or alcohol.

■ Calibration Cycle

It is recommended that the tester be calibrated once every year for correct operation;  
ask Yokogawa to do the periodic calibration for you.

■ Locking the MEAS Key (for Continuous Measurement)

The MEAS key, when pulled up to the right, can be locked to ensure the key remains turned on. 

Use this mechanism when making continuous measurement over a prolonged period. 

Note, however, that leaving the key turned on for an unreasonably long time will accelerate  
the discharge of the batteries.

■ Battery Life (reference only)

For MY10-03 at rated 500 V/100 MΩ: 

Approximately 10 hours when in continuously operation with central indication  

(approx. 2 MΩ; with standard supplied batteries).

Note

 

The data above is typical. Nevertheless, the battery life varies depending on the operating conditions. 
Check the batteries before measurement.

6.  Measuring AC Voltages

WARNING

 

• Do not press the MEAS key while measuring AC voltages. Otherwise, it may result in a failure.

• Voltage that exceeds the specified limit must not be applied to terminals.

6.1 Connecting the Earth Probe

• Securely connect the earth probe to the measured object’s 

ground line  

(if the measured object is not grounded, this process may be omitted).

6.2 Connecting the Line Probe

• Bring the line probe into contact with the measured object.

 

The pointer indicates the AC voltage. 

Read the value on the voltage measurement scale ( 

 V).

 

If the voltage of 20 V AC or more is present between the earth and 

 

line terminals, the ALARM LED turns on.

Note

 

When measuring insulation resistance, the tester can also be used to verify that there is no voltage present  
at the measured object.

7.  Using Protection Cover and Shoulder Strap

The tester comes with a protection cover and shoulder strap as standard accessories.

• The protection cover can be used as a front cover 

(for the indicator window)  

or as a bottom cover.  
(It is set as the front cover when 

delivered from the factory.)

• Using the shoulder strap allows you to position the tester in front of 

 

your chest for ease of reading.  

Pass the strap through the shoulder strap guide and adjust the length of 

 

the strap to allow you a good view of the tester.

• Remove the cover from the front, and attach it to the bottom using 

 

the fixing hole (B) on the surface of the cover. 

 

This is useful when the scale is too close to your body to see clearly  

(See the figure on the right).

• A belt on the cover which is fitted with pieces of Velcro, 

 

can be used to store the probes  

(Remove the probes from the tester terminals when storing them).

Protection cover 

fixing holes

Shoulder strap

Shoulder

strap guide

Protection

cover

A

B

8.  Battery Replacement

WARNING

 

• Remove the probes from the tester and then turn off the MEAS key before opening the casing to replace 

 

the batteries.

• Do not touch the MEAS key during replacement. Otherwise, a high voltage may be produced.

CAUTION

 

• Do not mix batteries of different types or new batteries with used ones.

• Always remove the batteries if the tester will not be used for a prolonged period of time. 

  If you store the tester with the batteries left installed, fluid is likely to leak from them, resulting in 

 

a malfunctioning of the instrument.

<Procedure>

1. Loosen the battery cover setscrew, and then slide the cover off of the main unit.

2. Replace all of the 4 batteries at the same time and make sure the polarities of 

 

the new batteries are exactly as shown on the battery holder.

3. After replacing the batteries, attach the battery cover and tighten the setscrew.

9.  Specifications

MY10-01

MY10-02

MY10-03

MY10-04

MY10-05

Rating

125 V/20 MΩ

250 V/50 MΩ

500 V/100 MΩ

500 V/1000 MΩ

1000 V/2000 MΩ

Central Indication

0.5 M

Ω

1 M

Ω

2 M

Ω

20 M

Ω

50 M

Ω

1st Effective

Measuring Range

0.02 to 10 M

Ω

0.05 to 20 M

Ω

0.1 to 50 M

Ω

1 to 500 M

Ω

2 to1000 M

Ω

2nd Effective

Measuring Range

0.01 to 0.02 M

Ω

10 to 20 M

Ω

0.01 to 0.05 M

Ω

20 to 50 M

Ω

0.05 to 0.1 M

Ω

50 to 100 M

Ω

0.5 to 1 M

Ω

500 to 1000 M

Ω

1 to 2 M

Ω

1000 to 2000 M

Ω

Lower Measuring 

Limit of Resistance

0.125 M

Ω

0.25 M

Ω

0.5 M

Ω

1 M

Ω

2 M

Ω

Rated Current

1 to 1.2 mA

1 to 1.2 mA

1 to 1.2 mA

0.5 to  0.6 mA

0.5 to 0.6 mA

AC Voltage

Measuring Range

0 to 250 V

0 to 300 V

0 to 500 V

0 to 500 V

0 to 500 V

<1>

<2>

<2>

Central indication

<First and second effective measuring ranges>

<1> 1st effective measuring range

<2> 2nd effective measuring range

The values which belong to both the first and 

second ranges in the table above, 

are included in the first range.

Item

Model

Standard test conditions

Ambient temperature and humidity:

23 ±5°C at 45 to 75% RH

Position:

Horizontal (within 5 degrees)

Influence of external magnetic field:

Earth magnetism

Battery Voltage:

Within effective range for the battery  
(When the battery is checked, the pointer must be within  

the BATT mark.)

Scale length:

Approx. 78 mm

Tolerances under the above conditions

Resistance measurement:

±5% of reading within the 1st effective measuring range

±10% of reading within the 2nd effective measuring range

Infinity and zero indications:

0.7% or less of scale length

AC voltage:

±5% of the maximum value

No-load voltage:

Within 130% of the rated voltage

Short-circuit current

12 mA or less

IM MY10-E <P2>