10-15
IM MV1000-01E
Computat
on and Report Funct
ons (/M1 and /PM1 opt
ons)
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
10.4 Startng/Stoppng Computaton
Dsplay
• Acton Performed When the START Key Is Pressed
Press
MENU
and then select
Menu
tab >
Math channel
>
Math start acton
.
Settngs
• Math start acton > Math start
Settng
Descrpton
Off
Does not start computation even when the START/STOP key is pressed.
Start
Start computation when the START/STOP key is pressed.
Reset+Start
Starts computation when the START/STOP key is pressed and resets the
computed result up to that point.
Procedure
• Startng Computaton
• Startng Computaton and Memory Samplng at the Same Tme
Press
START/STOP
. Computation starts at the same time as memory sampling.
The computation icon appears in the status display section.
* Math start must be set to Start or Reset+Start.
• Startng Computaton Only
1.
In Operation Mode, press
FUNC
.
The function menu appears.
2.
Press the
Math start
soft key.
Computation starts, and the status display section shows the computation icon.