5. SAFETY AND EMC STANDARDS
The following will be acquired.
Safety:
IEC/EN 61010-1 compliance (CE)
IEC/EN 61010-2-030 compliance (CE)
CAN/CSA-C22.2 No.61010-1 (CSA)
CAN/CSA-C22.2 No.61010-2-030 (CSA)
UL61010-1 (CSA NRTL/C)
UL61010-2-030 (CSA NRTL/C)
Overvoltage category: II
Pollution degree: 2
Measurement category: II
Rated power supply voltage: 15-30 V DC (±10%) 1.6 W or
100-240V AC/DC (±10%)
50/60 Hz 1.6 W 4.6 VA
Rated voltage (Input signal) : 0.5 V rms
Rated voltage (Analog output signal): −10 to +10 V DC,
0 to 20 mA DC
Rated transient overvoltage: 1500 V
CAUTION
This instrument is for Measurement Category II.
Do not use it for measurements in locations falling
under Measurement Categories III, and IV.
Internal Wiring
Outlet
Entrance
Cable
III
T
O
II
IV
Measurement
category
Description
Remarks
O
(other) For measurements
performed on circuits not
directly connected to MAINS.
II
CAT.II For measurements
performed on circuits directly
connected to the low-voltage
installation.
Appliances,
portable
equipments,
etc.
III
CAT.III For measurements
performed in the building
installation.
Distribution
board, circuit
breaker, etc.
IV
CAT.IV For measurements
performed at the source of
the low-voltage installation.
Overhead
wire, cable
systems, etc.
CAUTION
Precautions for using this product to comply with
safety standards:
The insulation class between the input and output of
this product is basic insulation (150V). When installing
this product, make sure that the insulation class of this
product satisfies the requirements for use. In addition,
supplementary insulation is required between the
output of this product and the connected equipment.
EMC standards:
•
CE:
EN 61326-1 Class A, Table 2
EN 61326-2-3
* The instrument continues to operate at a measurement
accuracy of within ±20% of the range during testing.
EN 55011 Class A, Group 1
EN 61000-3-2
EN 61000-3-3
•
RCM:
EN 55011 Class A, Group 1
•
KC:
Electromagnetic wave interference prevention standard,
electromagnetic wave protection standard
CAUTION
Caution to comply with EMC standards: When
operating this instrument by external power supply,
use an independent power unit conforming to CE-
marking. Be sure to use the lightning arrester to
comply EMC standards.
Note: When option code /C0 or /FB is specified, the conformity to
the safety and EMC standards is excluded.
6. ENVIRONMENT STANDARD
EU RoHS directive: EN IEC 63000
(However, when option code /C0 or /FB is specified, CE mark
-
ing is not applicable because the product does not comply
with the Safety and EMC standards.)
7. TRANSPORT AND STORAGE CONDITIONS
•
Temperature: −25 to 70°C
•
Temperature change rate: 20°C per hour or less
•
Humidity: 5 to 95%RH (no condensation)
CAUTION