3

IM 77J01B01-01E

6th Edition

Wiring Diagram

    

CTG-5

10

11

3

1

7

8

9

Power supply

11

10

8

L

+

N

GND

Input

+

9

7

Output

1

3

A

±

CTG-5

 CAUTION

 

Unused terminals should not be connected.

 

The power line and input/output signal lines should 

be installed away from noise-generating sources.  

Other wise accuracy cannot be guaranteed.

 

Make sure to earth ground the ground terminal 

through minimum resistance. The length and 

thickness of the grounding cable should be as 

short and thick as possible.  Directly connect the 

lead from the ground terminal (terminal no. 8) of 

the product to the ground.  Do not carry out daisy-

chained inter-ground terminal wiring.

 

The product is sensitive to static electricity; 

exercise care in operating it.  Before you operate 

the product, touch a nearby metal part to discharge 

static electricity.

 

If the ambient temperature is 50 °C or more, please 

use the cable that the rated temperature is 70 °C or 

more.

4.  MAINTENANCE

The product starts running immediately when the power is turned 

on; however, it needs 10 to 15 minutes of warm-up before it 

meets the specified performance.

4.1  Calibration Apparatus

 •

A AC voltage current standard (YOKOGAWA 2558A or the 

equivalent)

 • A digital multimater (YOKOGAWA DM7560 or the equivalent)

 •

A precision resistor of 250 Ω ± 0.01%, 1 W

4.2  Calibration Procedure

1. 

Connect the instruments as shown below.

    

CTG-5

10

11

3

1

7

8

9

Power supply

11

10

8

L

+

N

GND

1

3

Input

AC voltge 

current

standard

R: 250 

 precision resistor

for current output

DMM

7

9

Output

R

+

A

±

2.  Use the AC voltage current standard and apply input signals 

equivalent to 0, 25, 50, 75, and 100% of the input span to the 

product. 

3.  Check to see the corresponding output voltages are 0, 25, 50, 

75, and 100% respectively and within the specified accuracy 

rating. “R” is used for current output.

 • If the output signals are out of the accuracy rating range, 

adjust the output signal level using the zero and span 

adjustment volumes on front face of the product.

  

Zero-adjustment volume

Output

Span-adjustment volume