5-28
IM XL100C-E
• For the alarm values of p5 and p8, enter a
value using 5 digits or less, excluding the
decimal. For calculation channels (CA01 to
CA32), enter a value using 5 digits or less,
excluding the decimal.
SD
Sets the data and time
Syntax
SD p1,p2<terminator>
p1
Date (YY/MM/DD fixed form)
YY (year):
00
to
99
MM (month):
01
to
12
DD (day):
01
to
31
p2
Time (HH/MM/SS fixed form)
HH (hour):
00
to
23
MM (min):
00
to
59
SS (s):
00
to
59
Query
SD?
Example
Set the internal clock to 13:00:00, October 1,
2005.
SD 05/10/01,13:00:00
Description • The form of p1 and p2 is fixed to 8 characters.
Use the following form. Do not enter spaces
between the digits; otherwise an error will
occur.
p1 = YY/MM/DD (Lower two digits of the year/
month/day)
p2 = HH:MM:SS (Hour:minute:second)
• This command cannot be issued while logging
is in progress.
SW
Sets the sampling interval
Syntax
SW p1,p2<terminator>
p1
Sampling Interval
For 8 ch input:
100ms
,
200ms
,
500ms
,
1s
,
2s
,
5s
,
10s
,
20s
,
30s
,
1min
,
2min
,
5min
,
10min
,
20min
,
30min
,
60min
,
or 1h
For 16 ch input:
200ms
,
500ms
,
1s
,
2s
,
5s
,
10s
,
20s
,
30s
,
1min
,
2min
,
5min
,
10min
,
20min
,
30min
,
60min
,
or 1h
p2
Reserved
Query
SW?
Example
Sets the sampling interval to 10 s.
SW 10S
Description • This command cannot be issued while logging
is in progress.
• Parameter p2 is undefined.
ST
Sets the tag
Syntax
ST p1,p2<terminator>
p1
Channel number
Analog CH
8-CH input:
01
to
08
16-CH input:
01
to
16
Pulse CH
PL01
Logic CH
DI01
,
DI02
Calc. CH
CA01
to
CA32
Comm. CH
CO01
to
CO32
p2
Tag (up to 8 characters)
Query
ST[ p1]?
Example
Set the tag of channel 02 to TAG2.
ST 02,TAG2
Description • This command cannot be issued while logging
is in progress.
• For the characters that can be used for the
tags, see appendix 1, “ASCII Character
Codes.” Note that semicolons, commas,
backslashes, asterisks, and question marks
cannot be used.
SX
Sets the group
Syntax
SX p1,p2,p3<terminator>
p1
Group number
1
to
4
p2
Group name (up to 8 characters)
p3
Channel configuration (up to eight channels
from below)
Analog CH
8-CH input:
01
to
08
16-CH input:
01
to
16
Pulse CH
PL01
Logic CH
DI01
,
DI02
Calc. CH
CA01
to
CA32
Comm. CH
CO01
to
CO32
Query
SX[ p1]?
Example
Set channels 01, 03, 04 to 06 to group number 1
using a group name GROUP2. Display a blank
space between channels 01 and 03.
SX 1,GROUP2,01.OFF.03.04-06
Set the channel configuration by using periods to
separate each channel, a hyphen to specify a
range of channels, or OFF.
Description • This command cannot be issued while logging
is in progress.
• For the characters that can be used for the
group name, see appendix 1, “ASCII
Character Codes.” Note that semicolons,
commas, backslashes, asterisks, and
question marks cannot be used.
SL
Sets the alarm line
Syntax
SL p1,p2,p3,p4,p5<terminator>
p1
Reserved
p2
Reserved
p3
Alarm line display ON/OFF
ON
,
OFF
p4
Reserved
p5
Display Color
RED
,
GREEN
,
BLUE
,
B.VIOLET
,
BROWN
,
ORANGE
,
Y.GREEN
,
LIGHTBLUE
,
VIOLET
,
GRAY
,
LIME
,
CYAN
(blue green),
DARKBLUE
,
YELLOW
,
LIGHTGRAY
(olive),
PURPLE
Query
SL[ p1[,p2]]?
5.5 Setting Commands (Setting)