<7. Explanation of Basic Items>
7-7
IM 01S01C01-01EN
7.3.6 Other display settings
Setting display mode after backlight off (DISP_
QUIET_MODE)
This setting allows the user to set the display mode
after backlight off. DISP_QUIET_MODE: 0 = Stay
at last target (the display remains in the state it had
before backlight off), 1: Turn page cyclic (engages
scan mode after backlight off), 2: Display off (the
screen is turned off after backlight off).
Squawk (SQUAWK)
This function displays a notice that identifies the
communicating FVX110. Executing this function
alternates the screen shown in Figure 7.7. The
squawk display is automatically cancelled after
about a minute, but can also be cancelled by
turning the scroll knob.
F0707.ai
Figure 7.7 Screen displayed during squawk
operation