<2. Wiring and Installation>
8
IM 12A01A02-12E
9th Edition : Mar. 23, 2018-00
(A hole is drilled, if specified.)
FLXA21
FLXA202
(For sensor 2 cable)
For sensor 1 cable
For sensor 1 cable
(For sensor 2 cable)
For power supply
For power supply
For grounding cable
Figure 2.5
Cable gland diagram
CAUTION
Be careful not to be injured by the sharp hole edges on the housing.
Install the supplied cable gland as shown in Figure 2.6. When using an adapter for conduit work,
see Figure 2.7.
FLXA21
FLXA202
Cable gland
nut
O-ring
Gable gland
O-ring
Gable gland
Sleeve
(for grounding
cable line)
Cable gland
cap
Cable gland
nut
Gaskets
Washer
Cable gland
cap
Close up plug
(rubber)
Figure 2.6
Cable glands
The unused cable glands should be sealed with the supplied close up plug.
CAUTION
When installing cable glands, hold cable glands and tighten cable gland nuts to a torque of 6
N•m. If cable glands, not gland nuts, are tightened, O-rings may be come out from the proper
positions.