SD 12A01A02-02E 4/4
6th Edition: Sep. 2013 (YK)
P-
n
Wiring Diagram
*2
*4
*5
*6
*6
n n n n n n n
n n
n n n n n n n n n
+
–
*1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm.
*2: Connect the analyzer to gland. (Class D ground: 100 ohm or less)
The minimum cross sectional area of the protective grounding cable should be 0.75 mm
2
.
*3: This line is connected to a distributor or 24V DC power supply.
*4: Terminal numbers for each sensor module are shown below.
*5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive
conductivity or pH/ORP with the SENCOM sensor, only one module can be installed.
*6: The terminal box may be necessary depending on the sensor cable length and the distance between the
analyzer and the sensor.
The SENCOM sensor is to be connected directly to the analyzer without a terminal box.
*7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals.
*1
WTB10 or BA10
Terminal box
FLXA21 2-Wire Analyzer
Sensor
WTB10 or BA10
Terminal box
Sensor
PH Module
SC Module
ISC Module
DO Module
SENCOM Module
16
15
19
17
13
18
14
12
11
PH
NC
18
17
14
13
15
16
12
11
DO
15
16
14
17
13
12
11
ISC
NC
16
15
14
13
12
11
SC
87
86
NC
84
83
82
NC
SENCOM
Power supply
Output 1 (1-5V DC)
20 to 130V DC
or
80 to 138V AC, 47 to 63Hz
Power supply
20 to 130V DC
or
80 to 138V AC, 47 to 63Hz
HOLD
WASH
FAIL
Output 2 (1-5V DC)
Output 1 (1-5V DC)
Output 2 (1-5V DC)
Case of Distributor
PH201G (Style B)
Case of Distributor
SDBT
(*3)
*3
*7
*7
(*3)
A
+
– CMN CMN
B
C
D
F
H
L N
A
+
–
+
–
B
1
+
–
2
F
H
L N
b
a
d
c
f
e