3. Program functions

3.1 Automatic start of cleaning 

In principal, there are 3 different ways of starting 
an automatic cleaning cycle. They can also be 
combined in a useful manner.

Loop
 

A recurring cycle is run via the internal clock (loop). 
In doing so, after a parameterized measuring time, 
cleaning is started automatically, e.g. every four 
hours. After completion of the cleaning process, 
the retractable fitting moves to the measurement 
position and the cycle runs again.

 
Real Time Event
 

Automatic cleaning is started at a specific point in 

time which can be parametrized (Real Time Even), 
e.g. every day at 8.15 a.m., 12 noon and 4.30 
p.m.. After completion of the cleaning process, 
the retractable fitting moves to the measurement 
position and stays there until the next Real Time 
Event starts.

External Trigger

Cleaning is started via an external trigger. After 
completion of the cleaning and after opening the 
external trigger, the retractable fitting moves to the 
measurement position and stays there until the 
external trigger is closed again.

Loop + Trigger 
 

As described, cleaning is carried out in a defined 
cycle (loop). Further cleaning processes can 
additionally be started via an external trigger and 
the retractable fitting can be kept in the cleaning 
position. This function is used if the sensor is to be 
watered during operation and if the defined cycle 
is to be interrupted, or if the sensor is to be pulled 
back into the rinsing chamber for being protected 
while a strong agitator is running in the container.

Event + Trigger

As described, cleaning is carried out at defined 
points in time (Real Time Event). In addition, further 
cleaning processes can be started via an external 
trigger and the retractable fitting can be kept in the 
cleaning position. This function is used if the sensor 
is to be watered during operation and if the cycle 
is to be interrupted. Or if the sensor is to be pulled 
back into the rinsing chamber for being protected 
while a strong agitator is running in the container.

3.2 Seal water

For the short time during which the measuring 
window moves over the sealing elements, when 
the retractable fitting moves from the measurement 
position to the cleaning position or back, there is a 
connection between the measuring material and the 
rinsing chamber. To ensure that as little measuring 
material as possible can access the rinsing chamber 
and that the sealing elements are additionally rinsed 
during this time, a seal water function can be 
parameterized. 

Function
 

If the seal water function is active, the “Cleaning 
I” valve always opens before the retractable fitting 
is moved. This means that if the water pressure 
connected to the “Cleaning I” valve is higher than 
the process pressure, the water (seal water) runs 
through the rinsing chamber to the process during 
the movement of the retractable fitting and thus 
prevents measuring material from accessing the 
rinsing chamber. The sealing elements of the rinsing 
chamber are rinsed at the same time.

 The seal water function ensures a better cleaning 
of the sensor, maintenance of the sealing elements 
and increases the operating life of the sensor and of 
the sealings. The function should always be active, 
if possible.
 
 If you do not want seal water running to the 
measuring material or if this is not admitted, the seal 
water function can also be deactivated. The general 
function of the retractable fitting and the control is 
not impaired by that.

Summary of Contents for Exner RF20C

Page 1: ...IM 12B06K07 E E 1st Edition Instruction Manual RF20C EXmatic 460 Control unit...

Page 2: ...ery 13 4 1 Delivery scope 13 4 2 Delivery inspection 13 5 Assembly 15 5 1 Wall mounting 15 5 2 Electric connections 16 5 3 Pneumatic connections 17 5 4 Rinse valve connections 18 6 Parameterization 19...

Page 3: ...he connected retractable fitting is suited and able to function Carefully read the operating instructions of the used retractable fitting and observe the notes on safety contained in it Take appropria...

Page 4: ...is the cleaning times measuring intervals and the starting times can be parameterized and adjusted to the respective specification Input The control monitors the respective position indication of the...

Page 5: ...can be started via an external trigger e g from the pH transmitter The control unit RF20C EXmatic 460 is a completely independent control unit and can principally be operated without any connection to...

Page 6: ...Event In addition further cleaning processes can be started via an external trigger and the retractable fitting can be kept in the cleaning position This function is used if the sensor is to be water...

Page 7: ...0 300 seconds If this time is set to 0 this program step is left out completely By opening the Cleaning I valve a delivery pump can be triggered via an outlet trigger no 21 22 Pause If the sensor is...

Page 8: ...eep the control unit in the packaging There it is protected best until being assembled 4 2 Delivery inspection Before releasing the control unit for assembly you must ensure that packaging and unit ar...

Page 9: ...only use approved tools Mount the control cabinet as shown below Figure 6 Wall mounting CAUTION DANGER WARNING 15 5 1 Wandmontage Stellen Sie sicher dass gen gend Arbeitsraum f r den Betrieb der Armat...

Page 10: ...re 7 Connection diagram There is the risk for life and limb if the voltage supply is not disconnected Die Armaturensteuerung muss stromfrei geschaltet und die Druckluft drucklos sein Ist die Spannungs...

Page 11: ...4 Output 24VDC max 80mA Clean Pump II 31 Output 24VDC max 100mA measurement 32 Output 24VDC max 100mA measurement 33 Output 24VDC max 100mA service Hold 34 Output 24VDC max 100mA service Hold 35 Outpu...

Page 12: ...Manual Mode you reach the level for manual operation see chapter 6 3 When you select Parameters you reach the parameterization see chapter 6 4 When you select System you reach the system control see...

Page 13: ...n of the function please refer to chapter 3 6 4 Parameters You can parameterize the times and the functions of the control in the parameters menu If you have chosen Parameter in the main menu or durin...

Page 14: ...shows the currently set starting time You can change this value by pressing the soft keys Keep one soft key pressed for 2 seconds and then the value starts changing automatically By pressing the NEXT...

Page 15: ...you switch from hours to minutes and seconds and back By confirming with Enter you save the changed value By pressing the ESC button you leave the menu without saving Set Date In the system mode Set...

Page 16: ...display is shown Change Code Code Manual Mode Code Parameter Code System UP DOWN If you change or forget a pass code you block the system access Write the set values down and store them at a safe pla...

Page 17: ...Yearly inspections Check the compressed air connections for tightness Check the multi hose connections for tightness and for a stable fit at the fitting and the control valves Retighten the clamps in...

Page 18: ...external trigger 24V DC internal power supply for potential free contact Output for external relais Cleaning Pump I and Cleaning Pump II 24V DC 80mA max Output for status and alarm contacts 24V DC 100...