<6. PROCEDURES FOR CONVERTER OPERATION>
6-7
IM 12E04A02-02E
9th Edition : Mar. 16, 2018-00
6.2.4 Functions in <PROGRAM2> Mode
Table 6.6 gives details on the functions in the <PROGRAM2> mode.
Table 6.6 Functions in <PROGRAM2> Mode
F
Operation
Setting range/
selection
Initial
indication
Remarks
1 Setting of time range for automatic cleaning
0.1 to 24.0 hours
0002.0
2 Setting of cleaning calibration ratio
1 to 20 times
00001
3
Setting of number of cleaning
1 to 20 times
00005
4 Setting of cleaning time 1 *1
10 to 120 seconds
00030
5 Setting of cleaning time 2 *2
10 to 120 seconds
00030
6 Setting of drain time 1 *1
10 to 120 seconds
00010
7 Setting of drain time 2 *2
10 to 120 seconds
00010
8
Setting of filling time 1 *1
10 to 120 seconds
00100
9
Setting of filling time 2 *2
10 to 120 seconds
00100
A Setting of relaxation time
30 to 600 seconds
00150
B
C Setting of maintenance contact output
operation
_ _ 1/_ _ 2
_ _ 2/Closed
1: Open in operation
2: Closed in operation
D Setting of FAIL contact output operation
_ _ 1/_ _ 2
_ _ 1/Open
1: Open in operation
2: Closed in operation
E
Setting of high and low alarm contact output
operation
_ _ 1/_ _ 2
_ _ 1/Open
1: Open in operation
2: Closed in operation
F
Setting of operation of contact output during
automatic cleaning and zero-calibration
_ _ 1/_ _ 2
_ _ 2/Closed
1: Open in operation
2: Closed in operation
1.
Selection of execution/stopping of turbidity
rangeover detection function
ON/OFF
ON
ON: Execution
OFF: Stop
2.
Selection of execution/stopping of
disconnected-lamp detection function
ON/OFF
ON
ON: Execution
OFF: Stop
3. Selection of execution/stopping of lamp failure
detection function
ON/OFF
ON
ON: Execution
OFF: Stop
4.
Selection of execution/stopping of A/D circuit
failure detection function
ON/OFF
ON
ON: Execution
OFF: Stop
5.
Selection of execution/stopping of failure-
inmemory-comparison detection function
ON/OFF
ON
ON: Execution
OFF: Stop
6.
Selection of execution/stopping of reference
voltage failure detection function
ON/OFF
ON
ON: Execution
OFF: Stop
7.
Selection of execution/stopping of end-of-
lamp-life detection function
ON/OFF
ON
ON: Execution
OFF: Stop
8.
Selection of execution/stopping of automatic
calibration failure detection function
ON/OFF
ON
ON: Execution
OFF: Stop
9.
Selection of execution/stopping of high and
low alarm detection function
ON/OFF
ON
ON: Execution
OFF: Stop
A.
B.
C.
D.
E.
F.
*1:
Effective with Suffix code "-A2" (with automatic cleaning)
*2:
Effective with Suffix code "-A3" (with automatic cleaning and calibration)
Note 1: The values in the initial value column are those given when the memory storing the data is initialized. These may be partially
different from the values set upon shipment from the factory.
Note 2: If an unused function code which is selected, "_ _ _ _ _" is displayed.