<10. Appendix>
10-1
IM 12J05D02-01E
10. Appendix
Adjustable limits and defaults for settings in Commissioning Level
10.1 Setpoint
Variable
Default
Lower limit
Upper limit
*OX1
19.5 mg/L
0.0 mg/L
50.0 mg/L
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*OX2
1.0 mg/L
0.0 mg/L
50.0 mg/L
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*OX3
10.0 mg/L
0.0 mg/L
50.0 mg/L
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*OX4
10.0 mg/L
0.0 mg/L
50.0 mg/L
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*Sp.mA2
10.0 mg/L
0.0 mg/L
50.0 mg/L
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*T1
25.0 ºC/77 ºF
0.0 ºC/32 ºF
50 ºC/122 ºF
*T2
25.0 ºC/77 ºF
0.0 ºC/32 ºF
50 ºC/122 ºF
*T3
25.0 ºC/77 ºF
0.0 ºC/32 ºF
50 ºC/122 ºF
*T4
25.0 ºC/77 ºF
0.0 ºC/32 ºF
50 ºC/122 ºF
10.2 Range
Variable
Default
Lower limit
Upper limit
Range 1: Process 0% = 0.00
100% = 19.99
0.0 mg/L
0.0 ppm
0.0 %sat
50.0 mg/L
50.0 ppm
300 %sat
Range 2: Process 0% = 0.00
100% = 19.99
0.0 mg/L
0.0 ppm
0.0 %sat
50.0 mg/L
50.0 ppm
300 %sat
Temperature
0% = 0 ºC / 32 ºF
100% = 50 ºC / 122 ºF
0.0 ºC / 32 ºF
50 ºC / 122 ºF
Note: The range has a minimum span of: 1.0 mg/L, 1.0 ppm or 10 %sat.
In case of a temperature output 25 ºC or 50 ºF
10.3 Hold
Variable
Default
Lower limit
Upper limit
mA fix 1
10.00 mA
0 or 4 mA
20 mA
mA fix 2
10.00 mA
0 or 4 mA
20 mA
9th Edition: Aug. 05, 2015-00
Summary of Contents for EXA DO402G
Page 9: ...Blank Page...
Page 23: ...Blank Page...
Page 77: ...Blank Page...
Page 81: ...Blank Page...
Page 85: ...Blank Page...
Page 87: ...Blank Page...
Page 95: ...Blank Page...