Settings of the DX
Node Address Settings
Press
MENU
(to switch to setting mode), hold down
FUNC
for 3 s (to switch to basic
setting mode), and select the
Menu
tab >
Communication (PROFIBUS)
.
• Node Address
Set a node address in the range of 0 to 125.
Note
The node address of PROFIBUS-DP can be checked on the Network Information Screen of the
DX. You can open the Network Information Screen by pressing
FUNC
>
Network info
softkey.
Others
Status Output (option, /F1 and /F2)
If setting
Communication error
of the Status Output (/F1 or /F2 option) to
On
, a relay
output is provided when a PROFIBUS-DP communication error occurs within the DX. If
a communication error occurs, contact your nearest YOKOGAWA dealer for repairs.
See Section 2.9 of the
DX1000/DX1000N User's Manual (IM04L41B01-01E)
or
DX2000
User's Manual (IM04142B01-01E)
.
In Basic Setting Mode:
When the DX is in the basic setting mode, communications are available but input/output
data is invalid.
Connection to the PROFIBUS-DP Network