3-11
IM DR231-01E
Installation and Wiring
3
3.4 Connecting the Interface Cables
When connecting a personal computer to the instrument via a communication interface, observe
the following:
GP-IB
The GP-IB connector of the GP-IB communication module is a 24-pin connector of IEEE St'd
488-1978. Only use cables that conform to IEEE St'd 488-1978 as a communication cable.
Connection Procedure
Connect the cable as shown in the figure below.
GP-IB module
Ferrite core
Personal
computer
Ferrite core
Personal
computer
Ferrite core
Function
grounding
terminal
Function
grounding
terminal
DR130
DR231/DR241
•When connecting the cable, take note of the following.
• To reduce noise, use two ferrite cores (e.g., ZCAT 3035-1330 from TDK) at both ends of the
interface cable as shown above.
• Securely tighten the screws which fasten the GP-IB cable connector.
• If more than one equipment is connected, the same addresses cannot be assigned to different
equipment.
• Use only cables of 2 m or less to interconnect each equipment.
How to Assign an Address
The address can be assigned easily by turning the rotary dipswitch as shown in the figure below.
Any address can be set from "0" to "15"; the characters "A" to "F" on the dipswitch correspond to
the address "10" to "15" respectively.
0 12
3
4
5
678
9
F
E
D
C
B A
012
3
4
5
678
9
F
E
D
C
B A
Summary of Contents for DR130