10.5 Setting Burn-out/Reference Junction
Compensation
D I S P
M O D E
P R I N T
F U N C
I N S
D E L
R A N G E A L A R M
S E T
C H A R T
E N T E R
F E E D
M. F U N C 1
M. F U N C 2
RECORD
ALARM
CHART
KEYLOCK
R E C O R D
SETUP=BURN_OUT
↓
D ALARM A/D_INTG FILTER RELAY BURN_OUT
↑
>Select Setting Parameter
E S C
• Enter the SET UP menu in the following procedure:
1. When the display presents the operation display mode or a menu other than the SET UP
menu, first turn off the power switch.
2. Turn the switch back on while pressing the DISP key and continue pressing the DISP key
for about 5 seconds.
• Select/set
using the
keys.
• To escape from a lower menu, press the MODE (ESC) key. Its main menu will appear,
although new settings/selections will not be kept.
• New settings/selections will be kept when you reach the final display of that setting.
ENTER
ENTER
ESC
BURN OUT
RJC
ENTER
ENTER
ESC
ENTER
ESC
ENTER
001-01:BURN OUT=OFF
>Select Channel No.
001-01:BURN OUT=OFF
>Select Channel No.
001-01:BURN OUT=OFF
OFF DOWN UP
***BURN OUT SET***
001-01:RJC=INT
>Select Channel No.
001-01:RJC=INT
>Select Channel No.
001-01:RJC=INT
INT EXT
***RJC SET***
001-01:RJC=EXT
INT EXT
EXT VALUE=_ _ _ _ _0µV
>Ext Value Limit(– 20000~ 20000µV)
***RJC SET***