<8. Parameter Summary>
31
IM 01C22A01-01E
No.
Item
Description
Rewrita-
bility
Remarks
Default Value
Applicability
F
P
L
D23 DISP HRV
Engineering range, higher
range value
–19999 to 19999
As specified when
ordered.
D30 TEMP UNIT
Temperature setting units
deg C/deg F
deg C
D31 STAT. P. UNIT
Static pressure setting
units
Selected from mmH
2
O,
mmAq, mmWG, mmHg, Torr,
Pa, hPa, kPa, MPa, mbar,
bar, gf/cm
2
, kgf/cm
2
, inH
2
O,
inHg, ftH
2
O, psi, or atm
As specified when
ordered. If not
specified, MPa.
—
D40 REV OUTPUT
Output reversal
NORMAL/REVERSE
If not specified,
NORMAL.
D45 H/L SWAP
Impulse piping accessing
direction
NORMAL/REVERSE
NORMAL
—
—
D52 BURN OUT
CPU error
—
HIGH/LOW, –5 to 110%*
3
HIGH
D53 ERROR OUT
Hardware error
HOLD/HIGH/LOW, –5 to
110%*
3
HIGH
D60 SELF CHECK
Self-diagnostic messages
—
Same as A60
E
AUX SET 2
Auxiliary setting data 2
—
Menu name
E30 BI DIRE MODE
Bidirectional mode
OFF/ON
OFF
—
—
E60 SELF CHECK
Self-diagnostic messages
—
Same as A60
H
AUTO SET
Automatic setup
—
Menu name
H10 AUTO LRV
Automatic measurement
range lower range value
setup
–32000 to 32000
Displays the same
data as C21.
H11 AUTO HRV
Automatic measurement
range higher range value
setup
–32000 to 32000
Displays the same
data as C22.
H60 SELF CHECK
Self-diagnostic messages
—
Same as A60
J
ADJUST
Adjustment data
—
Menu name
J10 ZERO ADJ
Automatic zero adjustment
–5 to 110.0%*
3
J11 ZERO DEV.
Manual zero adjustment
J20 EXT. ZERO ADJ
External zero adjustment
screw permission
ENABLE/INHIBIT
J60 SELF CHECK
Self-diagnostic messages
—
Same as A60
K
TEST
Tests
—
Menu name
K10 OUTPUT in %
Test output % setting
–5 to 110.0%*
3
Displays
‘ACTIVE’ while executing
K60 SELF CHECK
Self-diagnostic messages
—
Same as A60
M
MEMO
Memo
—
Menu name
M10 MEMO 1
Memo
8 uppercase alphanumerics
M20 MEMO 2
Memo
8 uppercase alphanumerics
M30 MEMO 3
Memo
8 uppercase alphanumerics
M40 MEMO 4
Memo
8 uppercase alphanumerics
M50 MEMO 5
Memo
8 uppercase alphanumerics
M60 SELF CHECK
Self-diagnostic messages
—
Same as A60
P
RECORD
History of the errors
—
P10 ERROR REC 1
Last error
Display the error
P11 ERROR REC 2
One time before
Display the error
P12 ERROR REC 3
Two time before
Display the error
P13 ERROR REC 4
Three time before
Display the error
P60 SELF CHECK
Self-diagnostic messages
—
Same as A60
*1:
In case of Model EJA120A, static pressure cannot be measured. The display is always 0 MPa, but this is not a measured value.
*2:
When Optional code /F1 is specified, substitute the value with 0.1.
*3:
When Optional code /F1 is specified, substitute the value –5 with –2.5.
*4:
See MWP(max. working pressure) on the nameplate. B40 shows an approximate value of maximum pressure for the capsule.
P
a
ra
m
e
te
r S
u
m
m
a
ry
8