IM 1C22H1-01E
9-7
9. GENERAL SPECIFICATIONS
Item
Description
Code
T0908.EPS
Gold-plated diaphragm
T
Teflon film
(Note 3)
With FEP film and fluorinated oil
Working range: 20 to 150
°
C, 0 to 2 MPa {0 to 20 kgf/cm
2
} (Not usable under vacuum)
R
Operating temperature
correction
(Note 4)
Adjusting range: 80 to 300
°
C
V
Capillary without PVC
sheaths
When ambient temperature exceeds 100
°
C, or use of PVC is prohibited
C1
Failure alarm down-scale
(Note 5)
Output status at CPU failure and hardware error: –5%, 3.2 mA DC or less.
When combining with Optional code F1, output signal is –2.5%, 3.6 mA DC or less.
E1
A1
Stainless steel amplifier
housing
(Note 6)
Gold-plate
Amplifier housing material: JIS SCS14A stainless steel (equivalent to JIS SUS316 cast
stainless steel or ASTM CF-8M)
Update time: 0.125 sec or less, see GS for the response time
F1
Fast response
Oil-prohibited use
K5
Degrease cleansing and dehydrating treatment
Oil-prohibited use with
dehydrating treatment
Y
Sealant (liquided silicone rubber)is coated on surfaces of JIS SUS630 nuts used for cover
flange mounting.
Sealing treatment to JIS
SUS630 nuts
Painting
Lightning protector
K1
Degrease cleansing treatment
No serration
(Note 2)
Q
No serration work on the flange gasket surface (for ANSI flange only)
A
Transmitter power supply voltage: 10.5 to 32 V DC (9 to 32 V DC for Fieldbus
communication type) Allowable current: Max. 6000 A (1x40
µ
s),
Repeating 1000 A (1x40
µ
s) 100 times
P
h
X1
Amplifier cover only
D1
D3
P calibration (psi unit)
bar calibration (bar unit)
(See Table for Span and
Range Limits.)
Calibration units
(Note 1)
D4
M calibration (kgf/cm
2
unit)
Epoxy resin-baked coating
Color change
Coating change
Pressure test /Leak test
Certificate
M05
M06
M07
T31
T32
T33
T36
T37
T38
T39
Mill Certificate
(Test pressure)
2 MPa {20 kgf/cm
2
}
5 MPa {50 kgf/cm
2
}
10 MPa {100 kgf/cm
2
}
3 MPa {29.8 kgf/cm
2
}
7.7 MPa {77 kgf/cm
2
}
7 MPa {70 kgf/cm
2
}
14 MPa {140 kgf/cm
2
}
(Flange rating)
JIS 10K
JIS 20K
JIS 40K
ANSI/JPI Class 150
ANSI/JPI Class 300
ANSI/JPI Class 300
ANSI/JPI Class 600
(For model EJA118W)
(For model EJA118N)
Nitrogen (N
2
) Gas
Retention time: 10 minutes
Process flange, Block
Process flange, Block, Pipe, Base
High pressure side: Process flange, Block, Pipe, Base
Low pressure side: Process flange, Block
(For model EJA118Y)
Note 1: The unit of MWP (Max. working pressure) on the data plate of a housing is same unit as
specified by option code “D1, D3 and D4.”
Note 2: This item cannot be applied to Model EJA118W wetted parts material code H (Hastelloy C), T
(Tantalum), or U (Titanium). (In case for code H, T or U, serration work on the flange gasket
surface is not possible).
Note 3: Teflon film can only be specified for model EJA118W.
Note 4: Specify the process operating temperature for zero correction.
Example: Zero correction by process temperature 90
°
C.
Note 5: The hardware error indicates faulty amplifier or capsule. Standard output status (without /C1) is
up-scale of 110%, 21.6 mA DC or more.
Note 6: Applicable only for electrical connection code ‘2’, ‘3’ or ‘4’. Not applicable for optional code P
h
and X1. Not applicable for optional code JF1 and JS1.