5-83
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:CHANnel<x>:PROBe[:MODE]
Function
Sets or queries the probe attenuation setting of a
channel.
Syntax
:CHANnel<x>:PROBe[:MODE] {<NRf>|
C0_001|C0_002|C0_005|C0_01|C0_02|
C0_05|C0_1|C0_2|C0_5|C1|C2|C5|C10|
C20|C50|C100|C200|C500|C1000|C2000}
:CHANnel<x>:PROBe:MODE?
<x> = 1 to 8
<NRf> = See the DLM4000 User’s Manual for
this information.
Example
:CHANNEL1:PROBE:MODE 1
:CHANNEL1:PROBE:MODE?
-> :CHANNEL1:PROBE:MODE 1.000
:CHANnel<x>:PROBe:PZCalibrate
Function
Executes zero calibration of current probes for a
channel.
Syntax
:CHANnel<x>:PROBe:PZCalibrate
<x> = 1 to 8
Example
:CHANNEL1:PROBE:PZCALIBRATE
:CHANnel<x>:VARiable
Function
Sets or queries the voltage scale of a channel in
0.01 V/division steps.
Syntax
:CHANnel<x>:VARiable {<Voltage>|
<Current>}:CHANnel<x>:VARiable?
<x> = 1 to 8
<Voltage>, <Current> = See the DLM4000 User’s
Manual for this information.
Example
:CHANnel1:VARIABLE 1V
:CHANNEL1:VARIABLE?
-> :CHANnel1:VARIABLE 1.000E+00
:CHANnel<x>:VDIV
Function
Sets or queries the voltage scale (in V/division) of
a channel.
Syntax
:CHANnel<x>:VDIV {<Voltage>|
<Current>}
:CHANnel<x>:VDIV?
<x> = 1 to 8
<Voltage>, <Current> = See the DLM4000 User’s
Manual for this information.
Example
:CHANnel1:VDIV 2V
:CHANNEL1:VDIV?
-> :CHANnel1:VDIV 2.000E+00
5.6 CHANnel Group