IM CL255
17
AC Voltage V (Input impedance: 1MΩ)
(Crest factor (CF): 3.0 or less, peak voltage: 1200V or less)
Measuring Range (Auto-ranging)
Accuracy
0.50 to 39.99V
±1.5% rdg ±3dgt (50/60Hz)
±2.0% rdg ±4dgt (30 〜 1kHz)
15.0 to 399.9V
150 to 750V
Conversion method : AC coupled, true rms responding, calibrated to the rms
Auto-ranging
Lo : 0 to 39.99V (Shifts to Mid at 40.00V)
Mid : 15.0 to 399.9V (Shifts to Lo at 14.9V and to Hi at 400.0V)
Hi : 150 to 750V (Shifts to Mid at 149V. At 770V or above, "OL" is shown.)
Resistance Ω
Measuring Range (Auto-ranging)
Accuracy
0 to 399.9Ω
±1.5% rdg ±2dgt
150 to 3999Ω
Auto-ranging
Lo : 0 to 399.9Ω (Shifts to Hi at 400.0Ω)
Hi : 150 to 3999Ω (Shifts to Lo at 149Ω. At 4000Ω or above, "OL" is shown.)
Continuity Check
(Range fixed)
Measuring Range
Accuracy
10 to 399.9Ω
±1.5% rdg ±2dgt
The buzzer beeps when the resistance is 20Ω or less.
Frequency
Measuring Range
Accuracy
10 to 3999Hz
±1.5% rdg ±5dgt
OUTPUT (Output impedance: about 10kΩ)
Measuring Range
Output voltage (mVDC)
Accuracy
DC
0 to ±399.9A
0 to ±399.9mV
±1.5% rdg ±3mV
±150 to ±2000A ±15.0 to ±200.0mV
AC
0 to 399.9A
0 to 399.9mV
±1.5% rdg ±3mV(50/60Hz)
±3.0% rdg ±3mV(40 〜 1kHz)
150 to 1700A
15.0 to 170.0mV
1701 to 2000A
170.1 to 200.0mV
±3.5% rdg ±3mV (50/60Hz)