14
IM CL155
●
Power Source : Two R6P 1.5V battery
●
Battery Life : Approx. 80 hours (continuity)
●
Current Consumption : Approx. 7mA max. (Sleep function : Approx. 20
μ
A)
●
Sleep function : Automatically powered down in Approx. 10 minutes after the
last switch operation
●
Withstanding Voltage : 6300V AC, 50/60Hz for 5 sec. between electrical
circuit and housing case or metal part of the jaws
●
Insulation Resistance : 50M
Ω
or greater at 1000V between electrical circuit
and housing case or metal part of the jaws
●
Conductor Size : Approx. 54.5mm diameter max.
●
Dimensions : Approx.105(W) x 247(H) x 49(D) mm
●
Weight : Approx. 470g (with batteries)
●
Safety Standard: EN 61010-1, EN 61010-2-033
EN 61010-031, EN 61010-2-032
Measurement category IV AC/DC 300V, Measurement
category III AC/DC 600V,
Measurement category II AC/DC 1000V, Pollution
degree2, indoor use altitude 2000m or less
●
EMC Standard : EN 61326-1 Class B Table 1, EN 61326-2-2, EN 55011
Class B Group 1
●
Environmental standard: EU RoHS Directive compliant
For conformity to environmental regulations and/or
standards other than EU, contact your nearest
YOKOGAWA office (PIM113-01Z2).
●
Accessories : Test leads Model 98072
1set
R6P batteries
2
Carrying case Model 93034
1
User s
Manual
1
●
Optional Accessories :
Output cable Model 98076
Output cable (for terminal screw) Model 98077
6. Calibration and After-sales Service
Should any failure occur while you are using the tester, follow the instructions
given below. If the tester still fails to operate correctly and needs repair, contact
the vendor from whom you purchased the instrument or the nearest Yokogawa
Meters & Instruments sales office.
●
Turn off the POWER switch once, then turn it back on again.
●
If the tester does not turn on, replace the battery with a new one.
Calibration
It is recommended that the instrument be calibrated once every year.