ii

 

WARNING

Never make measurement on a circuit above 750V AC or 1000V DC.

 Do not use the instrument in an atmosphere where any flammable or 

explosive gas is present.

 Do not attempt to make measurement in the presence of flammable gasses, 

fumes, vapor or dust. Otherwise, the use of the instrument may cause 

sparking, which can lead to an explosion.

 Avoid using the instrument if it has been exposed to rain or moisture or if 

your hands are wet.

Do not exceed the maximum allowable input of any measurement range.

Never open the battery compartment cover when making measurement.

 Do not use the instrument if there is any damage to the casing or when the 

casing is removed.

 Do not turn the Function Selector switch with plugged in test leads con-

nected to the circuit under test.

 Do not install substitute parts or make any modification to the instrument. 

Return the instrument to Yokogawa Meter & Instruments or your distributor 

for repair or re-calibration.

 Always switch off the instrument before opening the battery compartment 

cover for battery replacement.  

  Do not use the test leads that have deteriorated or are detective.

  Check the test leads continuity

 

WARNING

To avoid damage to the instrument or electric shock!

The restrictions on the maximum voltage level for which the CL150 testers 

can be used, depend on the measurement categories specified by the safety 

standards. These category specifications are formulated to protect operators 

against transient impulse voltage in power lines.

Function

Maximum Allowable Input

MEASUREMENT

CATEGORY II

MEASUREMENT

CATEGORY III

 A

AC 2000A rms

Measuring circuit voltage : 

AC 750V rms

AC 2000A rms

Measuring circuit voltage : 

AC 600V rms

 V,   V

AC 750V rms/DC 1000V

AC 600V rms/DC 600V

Input terminal-to-

ground voltage

AC 750V rms/DC 1000V

Summary of Contents for CL150

Page 1: ...User s Manual CL150 Clamp on Tester クランプテスタ IM CL150 保証書付 IM CL150 2018 6 11版 KYOU ...

Page 2: ...e 4 2 3 Voltage Measurement 5 2 3 1 DC Voltage Measurement 5 2 3 2 AC Voltage Measurement 5 2 4 Resistance Measurement 6 2 4 1 Resistance Measurement Normal Mode 6 2 4 2 Continuity Check 400Ω range fixed 6 3 Other Functions 8 3 1 Sleep Function 8 3 2 Data Hold Function 8 3 3 OUTPUT Terminal For current measurement only 9 3 4 Optional Accessories 10 4 Battery Replacement 11 5 Specifications 12 6 Ca...

Page 3: ...o avoid the risk of personal injury or death and or damage to the instrument Double Insulation This symbol indicates double insulation AC Voltage Current This symbol indicates AC voltage or current DC Voltage Current This symbol indicates DC voltage or current Ground This symbol indicates ground earth This symbol Indicates that this instrument can clamp on bare con ductors when measuring a voltage...

Page 4: ...nstall substitute parts or make any modification to the instrument Return the instrument to Yokogawa Meter Instruments or your distributor for repair or re calibration Always switch off the instrument before opening the battery compartment cover for battery replacement Do not use the test leads that have deteriorated or are detective Check the test leads continuity WARNING To avoid damage to the i...

Page 5: ...priate terminal on the instrument Make sure to remove the test leads from the instrument before making current measurement Be sure to set the Function Selector switch to the OFF position after use When the instrument will not be in use for a long period of time Place it in storage after removing the battery Use a damp cloth and detergent for cleaning the instrument Do not use abrasives or solvents...

Page 6: ...ts ranges Also used to power the instrument on 4 Data Hold Button Freezes the display reading is shown on the display when Data Hold is enabled 5 Mode Selector Button On a AC current 400A or 2000A or the resistance range Press the mode button to cycle through the measurement modes The instrument is initially set to the normal mode and can be switched to peak or continuity check mode by means of th...

Page 7: ...C voltage output in proportion to the AC or DC current reading See section 3 3 OUTPUT Terminal The out put is connected to a recording device such as a chart recorder for long hour monitoring No output is available on voltage and resistance ranges 10 Lo Terminal Accepts the black test lead for voltage or resistance measurement 11 Hi Terminal Accepts the red test lead for voltage or resistance meas...

Page 8: ... before measurement 2 2 Current Measurement WARNING Do not make measurement on a circuit above 750VDC This may cause shock hazard or damage to the instrument or equipment under test Do not make current measurement with the test leads connected to the Hi and Lo terminals When measuring current is not less than 1000A make sure to stop meas urement within the maximum measuring time shown below Otherw...

Page 9: ...ws may buzz when measuring large current This has no effect on the instrument s performance or safety 2 2 2 Peak Current Measurement Peak Mode 1 Set the Function Selector switch to the 400A or 2000A position 2 Press the button once to enter from the normal mode to the PEAK mode should be shown on the display 3 Press the open close lever to open the transformer jaws and clamp them onto the conducto...

Page 10: ...ck test lead into Lo terminal 3 Connect the tip of the red and black test leads to the positive and negative sides of the circuit under test respectively Take the reading on the display 2 3 2 AC Voltage Measurement 1 Set the Function Selector switch to the V position AC should be shown on the upper left corner of the display 2 Slide the terminal cover to the left to disclose the Hi and Lo terminal...

Page 11: ... test leads to the circuit under test respectively Take the reading on the display NOTE When shorting the tip of the test leads the display may read a very small resistance instead of 0 This is the resistance of the test leads on the display 2 4 2 Continuity Check 400Ω range fixed The continuity check mode is enabled by pressing the button on resistance range and Ω is indicated on the display to s...

Page 12: ...t the tip of the test leads and check whether the display reads 0 5 Connect the tip of the test leads to the circuit under test If the resistance is 50 Ω or less the buzzer beeps NOTE When shorting the tip of the test leads the display may read a very small resistance instead of 0 This is the resistance of the test leads on the display ...

Page 13: ...s shown on the display This disables the sleep function and enables continuous use of the instru ment NOTE Connecting the plug to the OUTPUT terminal disables the Sleep function The function is enabled on removing the plug from the terminal The Sleep function is disabled in the PEAK measurement mode 3 2 Data Hold Function This is a function used to freeze the measured value on the display Press th...

Page 14: ...ad so that the output voltage can be connected to a recording device such as a chart recorder 2 lide the terminal cover to the right to disclose the OUTPUT terminal and insert the output plug into the terminal Make connection to the recording device 3 Set the Function Selector switch to the 400A or 2000A position and follow appropriate measurement instructions NOTE Output voltage is 1mV A on the 4...

Page 15: ...bility of a clamp meter With the use of the Clamp Adapter you can not only extend current range over 3000A but also clamp on a large bus bar or conductor 1 Set the Function Selector switch to the A position 2 As shown in the figure below clamp Model CL150 onto the pickup coil of Model 99025 3 Clamp Model 99025 onto the bus bar or conductor under test 4 Take the reading on Model CL150 and multiply ...

Page 16: ...ndicated in battery compartment If the battery voltage becomes too low for the instrument to operate normally is shown on the display Then replace the battery Note that when the battery is completely exhausted the display blanks without shown 1 Set the Function Selector switch to the OFF position 2 Unscrew and remove the battery compartment on the bottom of the instrument 3 Replace the battery obs...

Page 17: ... to 40 00V 0 01V 1 0 rdg 2dgt 50 60Hz 1 5 rdg 3dgt 40 to 1kHz 400V 15 0 to 400 0V 0 1V 750V 150 to 750V 1V Initially set to the 40V range Input impedance is about 1MΩ DC Voltage V Auto ranging Ranges Measuring range Resolution Accuracy 40V 0 to 40 00V 0 01V 1 0 rdg 2dgt 400V 15 0 to 400 0V 0 1V 1000V 150 to 1000V 1V Initially set to the 40V range Input impedance is about 1MΩ Resistance Auto rangin...

Page 18: ...perature and Humidity 0 to 40 relative humidity up to 85 without condensation Storage Temperature and Humidity 20 to 60 relative humidity up to 85 without condensation Effect of conductor position Within 2 0 rdg 3dgt to a 10 mm dia conductor at every part inside the jaws Effect of external magnetic field 1A or less in AC or DC magnetic field of 400 A m Power Source Two R6P 1 5V battery Battery Lif...

Page 19: ...ssories Output cable Model 98076 Output cable for terminal screw Model 98077 6 Calibration and After sales Service Should any failure occur while you are using the tester follow the instructions given below If the tester still fails to operate correctly and needs repair contact the vendor from whom you purchased the instrument or the nearest Yokogawa Meters Instruments sales office Turn off the PO...

Page 20: ...IM CL150 15 This manual is valid only in China 产品中有害物质的名称及含量 表示该有害物质在该部件所有均质材料中的含量均在GB T 26572 规定的限量要求以下 表示该有害物质至少在该部件的某一均质材料中的含量超出GB T 26572 规定的限量要求 有害物质 ...

Page 21: ......

Reviews: