OUTPUT (AC Current Ranges)
DC Output : 100.0mV per 1000 counts (Output impedance : about 10k
Ω
)
Ranges
Measuring range
Accuracy (Frequency Range)
400A
0 to 400.0mV/0 to 400A
±
1.5% rdg
±
0.5mV (50/60Hz)
±
2.5% rdg
±
0.5mV (40 to 1kHz)
2000A
0 to 150.0mV/0 to 1500A
±
1.5% rdg
±
0.5mV (50/60Hz)
±
3.5% rdg
±
0.5mV (40 to 1kHz)
150.0 to 200.0mV/1500 to 2000A
±
3.5% rdg (50/60Hz)
■ General Specifications
●
Operating System : Dual integration
●
Measurement Function : AC current, DC current, AC voltage, resistance,
continuity check
●
Display : Liquid crystal display with maximum counts of 4000
●
Overrange Indication : "OL" is shown on the display
●
Response Time : Approx. 2 seconds.
●
Temperature and Humidity for Guaranteed Accuracy : 23
℃
±
5
℃
, relative
humidity 45 to 75% without condensation
●
Operating Temperature and Humidity : 0 to 40
℃
, relative humidity up to
85% without condensation
●
Storage Temperature and Humidity : -20 to 60
℃
, relative humidity up to
85% without condensation
●
Effect of conductor position : Within
±
2.0%rdg,
±
3dgt to a 10 mm-dia
conductor, at every part inside the jaws
●
Effect of external magnetic field : 1A or less in AC or DC magnetic field of
400 A/m
●
Power Source : Two R6P 1.5V battery
●
Battery Life : Approx. 150 hours (continuity)
●
Current Consumption : Approx. 5mA max. (Sleep function : Approx. 20
μ
A)
●
Sleep function : Automatically powered down in Approx. 10 minutes after the
last switch operation
●
Withstanding Voltage : 6300V AC, 50/60Hz for 5 sec. between electrical
circuit and housing case or metal part of the jaws
●
Insulation Resistance : 10M
Ω
or greater at 1000V between electrical circuit
and housing case or metal part of the jaws
●
Conductor Size : Approx. 54.5mm diameter max.
●
Dimensions : Approx.105(W) x 247(H) x 49(D) mm
●