3.3 Resistance Measurement
Procedure
Setting the Function
1.
With the source value and measurement value displayed, press
FUNCTION 1
. The function
options are displayed.
2.
Use the arrow keys to select
Ω
. The display returns to the source and measurement value
display.
Set the measuring range.
3.
Under Function 1, press
RANGE
.
4.
Use the arrow keys to set the measurement range. The display returns to the source and
measurement value display.
Setting the 0% and 100% Values (when necessary)
5.
Set the 0% and 100% values according to section 3.7, “Setting the 0% and 100% Values”.
Setting the Wiring System
6.
With the source value and measurement value displayed, press
SETUP
under Function 1.
7.
User cursor keys to select
Connection Method
. Wiring systems appear in the selection menu.
8.
Use the arrow keys to set the wiring system.
9.
Press the arrow key corresponding to
EXIT SETUP
. The settings are confirmed, and the screen
reverts to show the source value and measurement value.
The cancel the settings, press
ESC
to return to the screen and step 6.
To initialize the settings, pressing the arrow key corresponding to
INIT SETUP
.