Setting and Executing Program Sweeps (CA550)
3.
Select
Program Sweep Setup
, and press
ENTER
.
CA550
Turn data save on or off.
Set the interval time.
Set the loop number of the calibration target.
Set the tag number of the calibration target.
Set the model of the calibration target.
Set the serial number of the calibration target.
* Output settings are on pages 2/3 and 3/3.
Setting the Interval Time
4.
Select
Interval Time
, and press
ENTER
. A list of options appears.
5.
Use the arrow keys to set the interval time, and then press
ENTER
.
Turning Data Save On and Off
6.
Select
Data Save
. ON and OFF appear in the selection menu.
7.
Use the arrow keys to set data save to ON or OFF.
When you set this to on, the measurement values and source values are saved automatically
when the sweep is completed.
Setting the Calibration Target Information (when necessary)
8.
Select
Model Number
, and press
ENTER
. An alphanumeric character input window appears.
9.
Enter the model number of the instrument to be calibrated.
Use the cursor keys to select characters, and then press ENTER. Press the arrow key
corresponding to DONE to confirm the entered character string.
For details on the alphanumeric input window, see chapter 3, “Common Operations,” Getting
Started Guide (IM CA500-02EN).
10.
Likewise, set information for
Serial Number
,
Tag Number
, and
Loop Number
.
2.9 Sweep Source
2-25
IM CA500-01EN