9-8
IM 755601-01E
9.5 Switching to the Talk-only Mode
Keys
NORMAL
MEAS&
LOCK
SAMPLE
EXTERNAL
REFERENCE
MANUAL
FAST
HI
HI LIMIT
EXIT
SETUP/MEAS
CHECK
MISC
I/F
MEAS
TIME
STORE
DATA
RECALL
DATA
LOCAL
ENTER
LIMIT
MODE
TRIG
MODE
SHIFT
R/%
REF
LIMIT
TRIG
7
1
0
.
2
3
5
4
6
8
9
LO LIMIT
IN
LO
M
Ω
M
Ω
k
Ω
Ω
,%
k
Ω
Ω
Ω
%
%
HIGH
SPEED
STORE
DATA
RECALL
DATA
CHECK
REMOTE
Procedure
The items that are specified or selected are confirmed when the ENTER (
Ω
,%) key is
pressed. When confirmed, the display returns to the top menu of the SETUP mode.
To exit from a menu in the middle of the operation, press the EXIT (R/%) key. The
display returns to the top menu of the SETUP mode.
Switching to the SETUP mode
1.
Press the
SHIFT
SETUP/MEAS
R/%
key to switch to the SETUP mode. If the “MEAS & LOCK”
indicator is off, this operation is not necessary.
2.
Set the device that is to receive the measured data to listen-only mode.
3.
5.
4.
7.
6.
9.
8.
(Mode)
(Add/not add the header)
SHIFT
REF
LIMIT
REF
LIMIT
REF
LIMIT
ENTER
Ω
,%
ENTER
Ω
,%
ENTER
Ω
,%
[Main display]
I/F
3
10. Press the
SHIFT
SETUP/MEAS
R/%
key to switch to the measurement mode. The “MEAS &
LOCK” indicator turns ON.
The measured data are output every time a measurement is made after the trigger
occurrence. The output data format is the same as the response to a “
:READ?
” query.
For details, see section 10.2.12, “Read Group.”
Explanation
Talk-only mode
Select “onLY” in the mode selection menu.
This mode is used to simply transmit data to other devices. It does not need to receive
a talker designation in order to transmit data.
It does not accept commands from the controller.