Phase Adjustment
Connect the probe to the CAL-output on the oscilloscope and turn the phase
adjustment screw with the included adjustment screwdriver to adjust the
observed waveform to a correct square waveform.
Under Compensated
Correct Waveform
Overcompensated
Maintenance
Cleaning
When cleaning the exterior of the probe, clean it using a soft cloth rag moistened
with either water or isopropyl alcohol. In this case, dry the probe completely
before starting the measurement.
Changing the Contact Tip
To change the contact tip, grip the contact tip firmly with pliers and carefully
pull it straight out of its contact socket along the axis of the probe. Do not grip
the white plastic insulator or the housing with pliers, because this will damage
the probe. If the contact tip is removed, the new contact tip can be inserted
with pliers into the socket, along the axis of the probe. To insert the contact
tip completely into the housing, press the contact tip against a hard surface
carefully.
Use pliers to grip and pull the contact
tip carefully out of its contact socket.
Do not grip the white plastic insulator
or the housing with pliers.