8
IM701941-01E
Handling
CAUTION
Since the spring type contact tip is very thin and sharp, great care
should be taken to prevent personal injury when the contact tip is
mounted. In particular, handle the spring type contact tip with great
care. The probe cable is a delicate part of the probe. If the probe
cable is bent or pulled forcibly, this may cause the cable to break.
To keep the accuracy and protect the product, do not apply any
impact or shock to the product.
Maintenance
Cleaning
When cleaning the exterior of the probe, clean it using a soft cloth rag moistened
with either water or isopropyl alcohol. In this case, dry the probe completely
before starting the measurement.
Changing the Probe Tip
To change the probe tip, grip the tip firmly with pliers and carefully pull it straight
out of its contact socket along the axis of the probe.
Adjustment Procedures
LF Compensation
Connect the probe to the CAL-output on the oscilloscope and adjust the “kHz”
trimmer (see also the Figs. next page.) so that the observation waveform
becomes the optimum square waveform.
Under Compensated
Correct Waveform
Overcompensated