background image

46

 

 

12.5

 

Menu

Configuration

Print

 

15/06/22  10:00 

PRINT 

Copy number    1 

 Printer       IR 
 Mode        fast 
 QR code       ON 
 Pairing BT 
 

KEY

 

FUNCTION

 

 

Enters  the  modification  mode  of  the 

selected data and then confirms it.

 

 

Selects the  available parameters.

 

In  modification  mode,  scrolls  the    

available values.

 

 

When  pressed  in  modify  mode  cancels 

the  selection  made,  otherwise  returns 

to the previous screen.

 

OK

 

>

 

>

 

ESC

 

PARAMETER

 

DESCRIPTION

 

Copy number

 

Sets the number of ticket copies to be printed.

 

This is a valid setting only if a printer has been selected.

 

Printer

 

Select the type of printer with which the ticket is printed:

 

BT

:

 

Bluetooth

®

  at  the  first  start  up  it  is  necessary  to  perform  the  paring 

 

procedure described below.

 

IR

:

 

Infrared.

 

OFF

:

 

None 

-

 the printer is turned off.

 

Mode

 

This parameter is visible only if the IR printer has been selected.

 

Selects the printing speed of the IR printer between 

fast

’ 

and 

slow

’.

 

Select  'slow'  in  order  to  make  the  printing  process  compatible  when  an  HP  IR 

printer is used.

 

QR code

 

QR code generation

 

ON

: pushing the button related to the interactive function 

Print

” 

the instrument 

generates  a  QR  code,  which  can  be  read  with  the 

YJ  Combustion

” 

APP  and 

allows  the  download  of  the  data  recorded  during  the  combustion  analysis  and 

additional measurements.

 

OFF

: the QR code will not be shown.

 

Pairing BT

 

 

Carry out the procedure to pair the Bluetooth

®

 

printer.

 

Summary of Contents for CA502

Page 1: ......

Page 2: ...4 8 Bluetooth connection 09 4 9 IR connection 09 4 10 Available software and applications 09 5 0 COMPONENTS DESCRIPTION 10 5 1 Instrument interface 10 6 0 TECHNICAL SPECIFICATIONS 12 6 1 Technical spe...

Page 3: ...o 39 12 2 6 Menu Configuration Analysis Air temp 40 12 3 Menu Configurationan Instrument 41 12 3 1 Menu Configuration Instrument Display 42 12 3 2 Menu Configuration Instrument Clock 43 12 3 3 Menu Co...

Page 4: ...cessories 63 19 3 Spare parts and servicing Service Centers 63 ANNEX A Analysis report examples 64 ANNEX B Optional measurements list 65 Measurement units matching abbreviations 66 ANNEX C Coefficient...

Page 5: ...MPORTANT INFORMATION Read information carefully and prepare safety appropriate action To prevent any danger from personnel or other goods Violating the information in this manual may cause danger to p...

Page 6: ...ofessional flue gas analysis for Small furnaces burning oil gas wood coal Low temperature and condensing boilers Gas heaters 2 3 Improper use of the product The CA502 should not be used As safety alar...

Page 7: ...on of a QR code with the purpose of downloading the data of the acquired measures having installed the YJ Combustion App which can be downloaded from the AppStore and PlayStore Measured values O2 CO C...

Page 8: ...intended to have a cleaning cycle of the pneumatic circuit The duration of the cycle depends on what has been set in the menu Configuration Analysis Autozero Used O2 sensors can be easily replaced by...

Page 9: ...on allows freedom of movement for the operator who is no longer tied directly to the instrument for acquisition and analysis 4 9 IR connection The CA502 analyzer is equipped with an infrared light int...

Page 10: ...5 6 7 3 5 0 COMPONENTS DESCRIPTION KEYS FUNCTION Activates the context keys shown on the display Turns on and off the instrument accesses the instrument menu If pressed for at least 2 seconds turns of...

Page 11: ...male connector of the gas temperature probe 4 IN pneumatic connector Input for the connection of the branch of the gas sampling probe with the condensation separating and anti dust filter assembly 6...

Page 12: ...ble 12 predefined by the factory Self diagnosis Checks all functions and internal sensors and reports any abnormal operation Temperature measurement Input for thermocouple type K with mignon connector...

Page 13: ...0 0 C 0 2 F 0 1 c 1 F 0 5 C 30 sec Flue gas temperature TcK sensor 4 1472 F 20 to 800 0 C 0 2 F 0 1 c 1 F 0 5 C 32 212 F 0 to 100 C 1 measured value 213 1472 F 101 to 800 C 30 sec Pressure draft diffe...

Page 14: ...e the instrument checked and calibrated once a year in order to comply with the existing standards 7 3 Analyzer power supply The CA502 contains a high capacity Li Ion rechargeable battery The battery...

Page 15: ...external power pack The CA502 can work with the batteries fully discharged by connecting the external battery charger provided 7 4 QR code generation The CA502 is able to generate and display a QR co...

Page 16: ...16 7 5 Connection diagram P N 68620 P N 68613 P N 68622 P N 68622 P N 68617...

Page 17: ...RFORM ANY MEASUREMENT WHEN THE FILTER IS REMOVED OR DIRTY TO AVOID ANY RISK OF IRREVERSIBLE DAMAGES TO SENSORS AND ANALYZER ITSELF 8 0 FLUE GAS ANALYSIS 8 1 FLUE GAS ANALYSIS 11 12 17 10 00 AUTOZERO P...

Page 18: ...e flue pipe For greater accuracy the user may insert the probe gradually into the pipe until the highest temperature is read The exhaust pipe must be inspected before carrying out the test to ensure t...

Page 19: ...int IN THIS SCREEN IT IS POSSIBLE TO MODIFY THE NUMBER OF THE MEMORY IN WHICH THE CURRENT ANALYSIS IS BEING SAVED SEE CHAPTER 11 0 11 12 17 10 00 PRINT Curr analysis Copy number 1 Printer OFF QR Code...

Page 20: ...Ta 15 4 C O2 4 2 CO 23 ppm CO2 2 9 COr p Pause 1 3 7 Automatically stores the first sample when the interval time is over SCAN THE QR CODE WITH THE YJ COMBUSTION APP IN ORDER TO DOWNLOAD THE RECORDED...

Page 21: ...e interval is over This condition is shown with the symbol When the Paused phase is over the interactive function Keep is shown By activating this function the recorded sample is stored to memory and...

Page 22: ...leaning When you finish using the sample probe clean it thoroughly as described below before returning it to its case Disconnect the sample probe from the instrument and from the water trap Fig a b th...

Page 23: ...igure below Clean all of the filter parts using only water then dry the components and reassemble the filter Replacing the particulate filter If the particulate filter appears black especially on the...

Page 24: ...a It can also show print or delete the recorded data and the additional measurements SEE CHAPTER 11 0 Configuration The user can set the different reference parameters of the instrument in order to pe...

Page 25: ...ty conditions of a specific working environment The instrument leaves our factory with the following pre set threshold values COmax 9 ppm Recommended exposure limit REL for 24 hr exposure in residenti...

Page 26: ...divided by a time gap defined by the user using the sub parameter interval It is possible to select the memory space to store the analysis and set the used fuel type In all modes the data displayed f...

Page 27: ...he measure will be printed on the ticket of the current combustion analysis Print Allows the option to print or generate the QR code for reading on the YJ Com bustion APP KEY FUNCTION Activate the key...

Page 28: ...ACTIVE OPERATION FUNCTION Save The measure will be printed on the ticket of the current combustion analysis Print Allows the option to print or generate the QR code for reading on the YJ Com bustion A...

Page 29: ...ng on the YJ Com bustion APP KEY FUNCTION Activate the keys shown on the bottom line of the display Returns to the previous screen Perform the Zero Sensor of the pres sure sensor ESC OK 15 06 18 10 00...

Page 30: ...ry 1 5 Status full Time 09 50 Date 11 12 17 Select KEY FUNCTION Activate the keys shown on the bottom line of the display Modifies the memory number and then confirms the changed setting When selectin...

Page 31: ...2 Tf 190 1 C Ta 15 4 C O2 4 2 CO 23 ppm CO2 2 9 COr p Print B 11 12 17 10 00 WARNING Printing in progress 11 12 17 10 00 PRINT Memory Copy number 1 Printer BT QR Code OFF Pairing BT Print 3 Print mem...

Page 32: ...ge analysis Analysis 1 Analysis 2 Analysis 3 Draft Pressure Delete Print Print Print 11 12 17 10 00 AVERAGE 1 2 Tf 190 1 C Ta 15 4 C O2 4 2 CO 23 ppm CO2 2 9 COr p Print 6 Detail of QR code generation...

Page 33: ...e measured gas goes above the set threshold If the alarm is in Off mode it is deactivated SEE CHAPTER 12 4 Print The print menu allows the settings for the print parameters The parameters include copy...

Page 34: ...ed from the CA502 If the relative humidity value is unknown the operator is recommended to enter 50 for the value SEE CHAPTER 12 2 2 O2 reference This mode allows the user the ability to set the oxyge...

Page 35: ...FUNCTION Returns to the previous screen Scrolls through the pages Confirms the choice of fuel to be used during the analysis ESC OK Example OK 11 12 17 10 00 FUEL 1 3 Natural gas Propane 2 Oil 4 Oil 6...

Page 36: ...air 0 0 15 06 18 10 00 CONDENSATION Altitude m R H air 0 100 Altitude above sea level Relative humidity of air KEY FUNCTION When pressed in modify mode cancels the selection made otherwise returns to...

Page 37: ...cancels the selection made otherwise returns to the previous screen The arrows select each line displayed In edit mode it scrolls through the suggested values Enters the modify mode for the selected p...

Page 38: ...ent unit can be set as ft m KEY FUNCTION When pressed in modify mode cancels the selection made otherwise returns to the previous screen The arrows select each line displayed In edit mode it scrolls t...

Page 39: ...the previous screen When in modify mode sets the desired value Enters edit mode of the selected element and then confirms the change Duration of autozero expressed in seconds Duration of the cleaning...

Page 40: ...When pressed in modify mode cancels the selection made otherwise returns to the previous screen OK ESC INTERACTIVE OPERATION DESCRIPTION T probe Measures the temperature by the Tc K probe connected to...

Page 41: ...user can select the date and hour format either in USA American or EU European mode SEE CHAPTER 12 3 2 Bluetooth In this sub menu it is possible to turn on and off the Bluetooth communication of the...

Page 42: ...otherwise returns to the previous screen The arrows select each line displayed In edit mode it scrolls through the suggested values Enters the modify mode for the selected parameter then confirms the...

Page 43: ...t 12h or 24h KEY FUNCTION When pressed in modify mode cancels the selection made otherwise returns to the previous screen Selects line in modification sets the value or the desired mode Enters the mod...

Page 44: ...the keys shown on the bottom line of the display Goes back to the previous screen ESC INTERACTIVE OPERATIONS DESCRIPTION Off Turns off Bluetooth On Turns on Bluetooth Bluetooth status On or Off Devic...

Page 45: ...e of the latter threshold Hysteresis Buzzer activation The measure blinks on the display The smoke suction pump DOES NOT turn off Type MAX alarm activation CO H2 NO Extract the probe from the stack Ma...

Page 46: ...e ticket is printed BT Bluetooth at the first start up it is necessary to perform the paring procedure described below IR Infrared OFF None the printer is turned off Mode This parameter is visible onl...

Page 47: ...art 15 06 18 10 00 PAIRING BT Printer detection in progress Restart B If the printer is not found the following message will appear No printer was detected Repeat the procedure pushing the button rela...

Page 48: ...turns to the previous screen Scrolls through the available languages Sets the selected language ESC OK 15 06 18 10 00 LANGUAGE 1 2 Espa ol English Fran ais 15 06 18 10 00 LANGUAGE 1 2 Espa ol English...

Page 49: ...15 06 18 10 00 RESTORE Clear memory data and restore factory setting Push OK to continue 15 06 18 10 00 RESTORE Deleting and resetting data KEY FUNCTION Exits the current screen without resetting Star...

Page 50: ...1 1 Gas probe Tests the tightness of the gas probe pneumatic path for potential leaks or possible damage SEE CHAPTER 13 1 2 Hardware If a malfunction occurs before contacting the Customer service col...

Page 51: ...n Selects the available parameters Shows the details about the selected sensor ESC Messages on the state and calibration of the electrochemical sensors Ok No problem detected absent The sensor was not...

Page 52: ...s of the gas sampling probe OK ESC Connect the flue gas sampling probe and filter unit assembly to the instrument Fully insert the black rubber cap 68624 on the gas probe tip as shown in the following...

Page 53: ...channels 11 12 22 10 00 VOLTAGES VBAT 3001 m VIN 4712 m VRTC 2602 m 11 12 22 10 00 HARDWARE Memories OK Calibration OK Voltages ADC channels 11 12 22 10 00 ADC CH 1 2 O2 GAS 1016 m CO SEN 58 m CO AUX...

Page 54: ...een next or previous screen ESC 09 05 22 10 00 INFO SERV 1 2 YELLOW JACKET Email custserv yellowjacket com 4266 E Street rd Trevose PA 19053 09 05 22 10 00 INFO SERV 2 2 Model CA502 S N 9999 FW rev 9...

Page 55: ...h as the reagents are consumed and when these have been used up completely the sensor must be replaced The sensors must be recalibrated on a regular basis to assure measuring accuracy recalibration ca...

Page 56: ...any condensate that may have formed Do not clean the instrument with abrasive cleaners thinners or other similar detergents 16 2 Preventive maintenance At least once a year send the instrument to the...

Page 57: ...ition of the sensor to be replaced be low is an example of a sensor to be replaced 3 Remove the tubes from the sensor to be replaced below is an example of a sensor with the disconnected pipes 4 After...

Page 58: ...table here below shows the minimum settling time for each sensor CODE DETECTED GAS POSITION SETTLING TIME O2 Sensor P N 68619 O2 Oxygen S1 8 hours CO Sensor with NOx filter 0 4000ppm CO Carbon Monoxid...

Page 59: ...e of the instrument 3 Remove the battery pack connector and replace the pack with a new one following the reverse procedure described above 2 Extract the battery pack Battery pack connector ONCE THE B...

Page 60: ...ntents of the zip file extension srec 3 Plug in the analyzer to the PC via the USB cable 4 Hold down the three circled buttons in above picture on the analyzer for at least 10 seconds 5 The display tu...

Page 61: ...rify the measured voltage values in Menu Diagnostic Hardware Voltages If VBAT 3000mV the battery needs to be changed If VIN 4700mV the output voltage of the battery charger is not sufficient to rechar...

Page 62: ...even start a Sample flow is obstructed Check that the water filter is clean and that it is not completely soaked Also check that the hose connected to the probe is not crushed b Sample intake flow is...

Page 63: ...Street Rd Trevose PA 19053 Email custserv yellowjacket com WWW YELLOWJACKET COM 19 2 Accessories PART NUMBER DESCRIPTION 68613 AC Power Adapter Kit For ALL Analyzers Power adapter w US plug adapter U...

Page 64: ...settings 11 12 22 10 00 ANALYSIS 1 5 O2 4 2 CO 23 ppm CO2 2 9 Tf 190 1 C Ta 15 4 C Es 91 4 Save Print Print Insert a customer email address Select subject matter line and fill out Add any job site no...

Page 65: ...be compared against minimum efficiency stated for the heating system performances c Ec LHV Condensation efficiency in relation to the Lower Heating Value Efficiency deriving from the condensation of w...

Page 66: ...66 Measurement units matching abbreviations ppm p mg m3 g mg Kwh w g GJ J ng J J g m3 G g Kwh W hPa h Pa P mbar b mmH2O H mmHg g inH2O i psi p C C F F m m ft ft...

Page 67: ...tere Calorifico Superiore Italian for HHV Higher Heating Value m H2O Mass of the air produced per each Kg of fuel in the combustion in stoichiometric condition m Air Mass of the air needed for combust...

Page 68: ...ed periods of time have to be delivered to YELLOW JACKET Laboratories postage paid The following cases are not covered by this warranty accidental breakage due to transport inappropriate use or use th...