YASUI koplight KS-1 Instructions For Use Manual Download Page 4

MD7-EUenAB-2110(8) 

Page 7/11 

2021-10-11 Rev. 8 

  

MD7-EUenAB-2110(8) 

Page 8/11 

2021-10-11 Rev. 8 

Troubleshooting

 

 

Problem 

Possible cause 

Solution 

LED does not illuminate 

LED switch button not pressed. 

Press LED switch button. 

Wrong battery polarity. 

Check battery polarity and set correctly 
(refer to “Assembly”). 

Batteries have run out. 

Replace with charged or new batteries. 

Tail cap not firmly tightened. 

Firmly tighten tail cap. 

Serial number label in battery compartment 
has peeled off and is blocking the battery 
terminals. 

Remove serial number label. 

Water has entered the battery compartment 
and caused a circuit malfunction. 

Contact the distributor. 

Cannot press the LED 
switch button 

Temperature too low. 

Warm the tail cap. 

Flickering/weak light 

Tail cap not firmly tightened. 

Firmly tighten tail cap. 

Batteries have run down. 

Replace with charged or new batteries. 

Short battery run time 

Electrical contact failures. 

Clean electrical contacts of the light 
handle, tail cap, and batteries using a dry 
cloth or cotton swab. 

Small battery capacity. 

Replace with larger-capacity batteries 
(NiMH rechargeable batteries of 

900 

mAh capacity are recommended). 

Product too hot 

Light handle may be damaged. 

Stop using the device immediately and 
contact the distributor. 

Contact your local distributor if the problem still exists after applying the solutions above. Repairs must be conducted by 
the authorized persons of the manufacturer. 

Appendices

 

 

Specifications

 

 

  Product name: koplight blade 

Model 
code 

Tip width 
(mm)

 

Blade length 
(mm) 

Total length 
(mm) 

KS-1 

10 

30 

139 

KS-2 

13 

45 

139 

KS-3 

18 

45 

139 

KS-4 

18 

59 

113 

KS-5 

25 

84 

113 

Material: Polycarbonate

 

 

  Product name: koplight HT blade 

Model 
code 

Tip width 
(mm) 

Blade 
length 
(mm) 

Total length 
(mm) 

KS-4H 

18 

61 

122 

KS-5H 

25 

86 

122 

KS-6H 

25 

111 

122 

Material: Polycarbonate

 

 

  Product name: koplight light handle

 

Model code 

KG-1 

Material 

Body: 

Anodized aluminum 

Lens: 

Polycarbonate 

LED switch button:  Fluorine resin 

Battery

 

2 × AAA nickel-metal hydride (NiMH) rechargeable battery 

Recommended battery 

Manufacturer: 

FDK CORPORATION 

Model code: 

HR-4UTHC 

Nominal voltage: 

1.2 V 

Battery capacity: 

900 mAh 

Rated input 

2.4 V DC 

LED 

Color temperature: 5,000 K 
Note: There are slight variances in the color temperature and brightness of the LED. 

Dimension 

Diameter: 15 mm (narrowest point) × Total length: 155 mm 

Operating/storage 
temperature 

0 °C to 40 °C 
Note: Recommended operating temperature range is 15 °C to 40 °C. Waterproof 
performance degrades in between 0 °C to 15 °C due to hardened O-ring. 
Recommended to store at room temperature. 

Relative humidity 

30% to 90%, non-condensing 

Atmospheric pressure 

70 kPa to 106 kPa 

Storage/Transfer 
condition 

Remove batteries 

  Accessories

 

Product name 

Model 
code 

Material 

koplight light handle finger hook 

KY-1 

Polycarbonate 

koplight light handle O-ring 

KP-1 

Fluorine resin 

koplight battery inserter 

KD-1 

Polycarbonate 

Retail packaging

 

 

Product name 

Content 

koplight blade/koplight HT blade 

12 × blades per package. 

koplight light handle 

1 × light handle, 1 × finger hook, 1 × O-ring, 1 × battery inserter per package. 
Note: Batteries are not included. 

Safety classifications 

 

Protection against electrical shock 

Internally powered ME equipment 

Degree of protection against electric shock 

Blade is type BF applied part 

Method of sterilization or disinfection 

Refer to "Cleaning and disinfection" and "Sterilization" 

Degree of protection against ingress of water 
(IEC 60529) 

IPX7 

This product is not for use in an oxygen rich environment. 

Precautions for

 

interference with other equipment 

 

In order to avoid LED lighting malfunction caused by interference with other equipment: 

  Keep away from portable wireless communication devices such as cellular phones by more than 30 cm. 

  Do not use close to or stacked on other equipment. 

  If electrosurgical instruments can cause malfunctions, try not to use them at the same time. 

Note: No malfunction will occur when the device comes into contact with an electric scalpel. Electrosurgical 
instruments are generally compatible with the device, but some may cause malfunctions. For further information, refer 
to “Electromagnetic Compatibility (EMC).”  

Electromagnetic Compatibility (EMC)

 

 

This product complies with EMC Standard EN60601-1-2: 2015. This product is intended for use in the electromagnetic 
environment specified in the following tables. The following problems may be encountered if the product is used in an 
unintended environment: 

  LED light does not turn on. 

  LED light turns on, but flickering occurs. 

  LED light turns on, but there is a significant increase or decrease in illumination that can be visually confirmed. 

  LED switch button does not work. 

 