YASUI koplight KS-1 Instructions For Use Manual Download Page 3

MD7-EUenAB-2110(8) 

Page 5/11 

2021-10-11 Rev. 8 

  

MD7-EUenAB-2110(8) 

Page 6/11 

2021-10-11 Rev. 8 

7.  Rinse 

3 minutes 

Purified water 

Cold 

N/A 

8.  Thermal disinfection

 b)

 

2.5 minutes 

Purified water 

93 °C 

N/A 

9.  Dry 

Until 
thoroughly dry 

N/A 

Minimum 
100 °C 

N/A 

a)  Washer-disinfector should be compliant with EN ISO 15883 series. 
b)  Disinfection process may be required depending on the guidelines of your country. This product is compatible with 

thermal disinfection at temperatures between 80 °C to 93 °C. 

Inspection and maintenance

 

 

Inspection

 

 

Inspect the items below before each use. Do not use the device if any problems are found. 

  No damage or deformation 

  No dirt, foreign object, and/or peel-offs on surface 

  No cracks or major scratches on blade 

  No cracks on lens 

  Light handle’s head (LED side) does not rotate 

  Tail cap can be firmly tightened 

  LED switch button is not broken and LED works 

Maintenance 

 

  Wipe the metal wear debris off when it is found around the tail cap’s screw thread. It is 

recommended that you apply lubricant: 
1.  Remove tail cap. 
2.  Apply suitable amount of medical lubricant to a cloth. 
3.  Hold light handle and point the battery compartment hole downward. 
4.  Wipe screw thread with the cloth, paying attention that the lubricant does not enter the 

battery compartment. 

Note: Do not directly apply lubricant to the light handle. 

  Damaged O-ring should be replaced to keep the light handle waterproof: 

1.  Unfasten tail cap.

 

2.  Remove damaged O-ring.

 

3.  Set new O-ring in the right position (figure 12), paying attention not to scratch it.

 

4.  Check that new O-ring is not distorted.

 

5.  Attach finger hook (optional), and firmly tighten the tail cap. 

(fig. 12) 

Packaging

 

 

  The device should be packaged in sterile packaging before sterilization. The 

packaging material and procedure should be in accordance with EN ISO 
11607-1 and EN ISO 11607-2. 

  Prepare light handle as follows before packaging. Noncompliance can lead to 

battery explosion during sterilization and/or incomplete sterilization: 
1.  Check that there are no batteries in the battery compartment. 
2.  Firmly tighten the tail cap. 
3.  Loosen the tail cap by one (1) rotation (about 360°, figure 13). 
Note: Do not completely separate the tail cap during sterilization unless you 
use ethylene oxide gas sterilization. Doing so can cause malfunction. 

Sterilization

 

 

 

Precautions for sterilization

 

 

  There will be malfunctions and/or material degradation when the device is sterilized by any methods other than the 

recommended methods. 

  Do not change the sterilization method during the product lifetime. 

Recommended sterilization

 

 

  Applicable sterilization methods 

Product 

Steam 

Hydrogen peroxide 

Ethylene oxide gas 

Light handle 

Not applicable 

Applicable 

Applicable 

Blade, finger hook, and 
battery inserter 

Applicable 

Applicable 

Applicable 

  Steam sterilization parameters 

Item 

Recommended condition 

Precautions 

Cycle type 

Pre-vacuum 

  Steam sterilization must not be applied to the light 

handle. 

  Temperature must not exceed 121 °C. 

Temperature 

121 °C 

Exposure time 

30 minutes 

  Hydrogen peroxide sterilization parameters 

Item 

Recommended condition 

Precautions 

Temperature 

47 °C to 55 °C 

  Repeated hydrogen peroxide sterilization may cause 

early material degradation.

 

Exposure time 

47 minutes 

Note: The manufacturer used a sterilizer equivalent to STERRAD®100NX® in the validation. 

  Ethylene oxide gas sterilization parameters 

Item 

Recommended condition 

Precautions 

Temperature 

50 °C to 60 °C 

 

Relative humidity 

40% 

Gas concentration 

300 to 1100 mg/L 

Exposure time 

4 hours 

Aeration time 

8 hours 

Storage

 

 

  Check for the points below before storage: 

o

  Batteries are removed 

o

  Device is cleaned 

o

  Device is thoroughly dried, including battery 

compartment 

o

  Tail cap is firmly tightened 

  Store the product under the conditions below: 

o

  Clean place at room temperature 

o

  No direct sunlight 

o

  Avoid ultraviolet light 

o

  Avoid high humidity 

 

Disposal

 

 

Dispose of the product according to the institute procedures and any applicable laws, regulations, and rules of your 
country.

  

Limitation of liability/Incident reporting

 

 

Yasui Co., Ltd. does not assume any liability when the product is misused and/or mishandled. Read these Instructions 
for Use carefully before use and keep in an easily accessible place for later reference. Only professionally trained 
healthcare personnel are allowed to use this product. 

If a serious incident occurs in relation to this product, it should be reported to the manufacturer and/or distributor. In the 
European Union, it should also be reported to the competent authority of the member state in which the user and/or 
patient is located. 

 

O-ring 

With O-ring 

O-ring 
groove 

Without O-ring 

(fig. 13 – loosen 360°)

 

Tail cap           LED switch button

 