39
Specification
Model
SI401
SI601
Category / General name
Instrument 3 Medical Sterilizer / Drying Sterilizer
Temperature
adjustment accuracy
2deg (Middle of furnace, Set temp.: 260
℃
)
Temperature
distribution accuracy
20deg (Set temp.: 260
℃
)
Temperature rise time
Approx. 70min. (Room temp. to 260
℃
)
Temperature fall time
Approx. 200min. (260 to 50
℃
)
Heater
Iron-chrome heater 1.2kw
Iron-chrome heater 1.36kw
Controller
VS4 type
Temperature control
system
PID control for heater output by microcomputer
Setting system
Digital setting by menu key and up/down keys
Operation mode
Fixed temperature, Quick auto stop, Auto stop, Auto start, Program operation
Sensor
K-thermocouple
Additional functions
Lock function, Auto recovering after power failure, Calibration offset
Self-diagnostic
functions
Failure of Sensor, heater, SSR, memory, internal communication,
temperature inputting circuit, automatic overheating prevention device,
overheating prevention device, measurement temperature
Safety device
Earth leakage breaker, Overheating prevention device
External dimensions
(W
×
D
×
H mm)
550×540×777
700×640×877
Internal dimensions
(W
×
D
×
H mm)
450×430×400
600×530×500
Capacity
77L
159L
Door
Single door, silicon packing
Power supply
(50/60Hz)
115V AC 10.6A
115V AC 11.8A
Weight
Approx. 42Kg
Approx. 59Kg
Accessories
2 Shelves (withstand load: 15kg/one shelf), Instruction manual
The performance under the power supply condition of AC 115V are shown here.
The usable ambient temperature of the unit is from 5
℃
to 35
℃
.
The temperature fall time here shows the reference value.