Yamato SI 401 Instruction Manual Download Page 16

 

 

13 

 

Operation Method 

Operation Mode and Function List 

 

The operation function of this unit is as follows; 

No. 

Name 

Description 

Page 

1. 

Overheating 
  prevention 
  function 

Auto overheating 
  prevention function 

This  function  is  set  to be  automatically activated 
(auto  reset)  when  the  temperature  exceeds  the 
setting temperature by 12

15 

Overheating 
  prevention device 

Though  the  device  shares  power  source,  display, 
and  key  input  with  the  controller,  it  has 
independent  temperature  measurement  circuit, 
CPU,  sensor  and  output  circuit.    Overheating 
prevention  temperature  can  be  set  using  the 
operation panel. 
The  unit  stops  operation  when  the  device  is 
activated.    The  unit starts  operation  again  when 
the  POWER  switch  is  pressed  again  (manual 
reset). 

2. 

Calibration offset function 

This  calibration  offset  function  is  for  calibrating 
the  difference  occurred  between  the  required  in- 
furnace  temperature  and  control  temperature 
(sensor  temperature)  of  the  controller.  This  unit 
can be calibrated toward either plus side or minus 
side of the whole temperature range. 

31 

3. 

Overheating prevention 
  temperature calibration 
  function 

The temperature of overheating prevention device 
is  automatically  corrected  when  the  temperature 
of controller is collected. 

4. 

Recovery at power failure 

The unit starts operation with the same condition 
as  just  before  power  failure  if  it  occurs  during 
operation.    Press  the  START/STOP  key  to  start 
the unit again. 

5. 

Setting value locking 

This  function  locks  the  established  operation 
status. 
It  can  be  set  and  cancelled  with  the  SUBMENU 
key. 

31 

 

Summary of Contents for SI 401

Page 1: ...nk you for purchasing Drying Sterilizer SH Series of Yamato Scientific Co Ltd To use this unit properly read this Instruction Manual thoroughly before using this unit Keep this instruction manual arou...

Page 2: ......

Page 3: ...Setting of Overheating Prevention Device 15 Sterilizing Operation 16 Fixed Temperature Operation 18 Quick Auto Stop Operation 19 Auto Stop Operation 20 Auto Start Operation 22 Program Operation 24 Oth...

Page 4: ...a problem that may cause a serious accident or ev en f atality CAUTION If the caution is ignored there is the danger of a problem that may cause injury damage to property or the unit itself Meaning o...

Page 5: ...ution electrical shock Caution scald Caution no road heating Caution not to drench Caution water only Caution deadly poison Prohibit Prohibit generally Prohibit inflammable Prohibit to disassemble Pro...

Page 6: ...it is dangerous and not recommended Do not use the power cord if it is bundled or tangled Do not use the power cord if it is bundled or tangled If it is used in this manner it can overheat and fire m...

Page 7: ...lectric shock may be caused Do not use a branching receptacle which may cause the heat generation 2 Choose a proper place for installation Do not install this unit in a place where Rough or dirty surf...

Page 8: ...Never use this unit in an area where there is flammable or explosive gas This unit is not explosion proof An arc may be generated when the power switch is turned ON or OFF and fire explosion may resul...

Page 9: ...xpectible trouble or malfunction CAUTION 6 Do not make an overload 7 Do not set samples in close formation The withstand load of shelf is 15kg uniform load Set the samples apart each other The tempera...

Page 10: ...on To prevent take measures that putting up a notice of operating etc 10 Handling of power code Two shelves are attached on the product One of them is previously fixed on the lowermost stand of bracke...

Page 11: ...he cord s insulation may melt and cause a fire or electrical shock If the power cord becomes damaged wiring exposed breakage etc immediately turn off the power at the rear of this unit and shut off th...

Page 12: ...9 Description and Function of Each Part Main Unit Temperature sensor Main switch Earth leakage breaker Control panel Door Door handle Power code Exhausted opening SI401 601 Front view Rear view...

Page 13: ...TO STOP Lamp Blinks while setting quick auto stop timer or auto stop timer Lights while quick auto stop timer or auto stop timer is running AUTO START Lamp Blinks while setting auto start timer Lights...

Page 14: ...gram Operation Sv 1 Program Temperature Setting Used for setting the temperature for each step in the program Sv 1 to Sv 30 is shown t 1 Program Time Setting Used for setting the time for each step in...

Page 15: ...ter specified period 19 4 Auto Stop Operation This operation is used to specify the automatic stop time in the fixed temperature operation Pressing the TIMER key displays AS p The setting temperature...

Page 16: ...again when the POWER switch is pressed again manual reset 2 Calibration offset function This calibration offset function is for calibrating the difference occurred between the required in furnace temp...

Page 17: ...14 Operation Method Operation Mode Function Setting Key and Characters The operation mode setting and function setting use the key operation and characters show in the following figure...

Page 18: ...e current temperature in furnace operation mode character and setting temperature of overheating prevention device are displayed on respective screens 2 Set the temperature for overheating prevention...

Page 19: ...ation on the center display screen 3 Start sterilizing operation Press the orange START STOP key for one second The operation starts and the PROGRAM lamp lights on After one hour sterilizing operation...

Page 20: ...n displays t 1 and the setting temperature screen displays 0 01 Do not change the value Press the ENTER key again The measurement temperature screen displays SV 2 and the setting temperature screen di...

Page 21: ...setting Also it displays the current setting temperature with blinking The FIXED TEMP lamp blinks too Set the temperature by pressing the 4 Start operation Press the orange START STOP key for about o...

Page 22: ...ustment is needed 2 Start timer operation Press the START STOP key for one second after deciding the time Timer operation starts with the FIXED TEMP and AUTO STOP lamps lighting on The timer is activa...

Page 23: ...the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting temperature display screen displays the current setting time wi...

Page 24: ...h the FIXED TEMP and AUTO STOP lamps lighting on Press the START STOP key to terminate the timer operation mode The screen returns to the initial setting screen To correct or check setting Changing th...

Page 25: ...TO START lamp blinks too Set the temperature using the Press the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting tem...

Page 26: ...heck setting Changing the setting temperature or time during operation is possible by pressing the TIMER key Use the to change the setting value Press the ENTER key respectively after changing the set...

Page 27: ...apability of increasing or decreasing temperature by 50 within 15 minutes about 30 minutes is needed to increase or decrease temperature by 100 from current temperature Repeat function Repeat function...

Page 28: ...program mode program pattern Press the PROGRAM key once The measurement temperature display screen displays the previous program mode Press the PROGRAM key again to display the next program mode PrG1...

Page 29: ...fall capability of unit For example about 75 minutes is needed to increase the temperature from room temperature to 250 for SH401 type 80 minutes for SH601 type When the current temperature is suppos...

Page 30: ...e set in increments of a minute under 99 hours and 59 minutes It can be set in increment of ten minutes over 100 hours The can change the setting time quickly when it is pressed continuously Press the...

Page 31: ...mperature screen indicates PS 7 in the example because repeat function is used at the seventh step The step number 1 to 7 can be input in the setting temperature display screen Enter the number 1 in t...

Page 32: ...n Method Program Operation Programming Preparation Form 1 Please use this form by making copies Register with PrG1 PrG2 PrG3 PAt1 PAt2 PAt3 No Project Name Date Programmer Program Pattern 250 200 150...

Page 33: ...PAt1 PAt2 PAt3 No Project Name Date Programmer Input Value Temperature Time min Repeat Function Step 1 To Times Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Step 8 Step 9 Step 10 Step 11 Step 12 Step 13...

Page 34: ...ature display screen falls by the setting temperature while the temperature on furnace rises When it is set to the minus side the temperature on the measurement temperature display screen rises by the...

Page 35: ...storm During a thunderstorm turn off the power key immediately then turn off the circuit breaker and the main power If this procedure is not followed fire or electrical shock may be caused When open c...

Page 36: ...cause trouble Never put sample on the base Fix the shelf on the bracket then set sample on it Recovering after power failure When power is supplied after a power failure the device automatically star...

Page 37: ...ntenance after returning the temperature of this unit to the normal one Do not disassemble this unit CAUTION Use a well drained soft cloth to wipe dirt on this unit Do not use benzene thinner or clean...

Page 38: ...equest you to disassemble this unit as possible and recycle the reusable parts considering to the environmental protection The feature components of this unit and materials used are listed below Compo...

Page 39: ...ort circuit detection ALARM lamp lights on Er 02 appears Triac is in short circuit Make a call for service Heater disconnecting detection ALARM lamp lights on Er 03 appears Heater is disconnected Make...

Page 40: ...ature fluctuates during the operation Too much samples Wind from air conditioner directly blows The change of ambient temperature is remarkable Samples are too moist The power supply voltage is lower...

Page 41: ...les office Check following items before contact Model Name of Product Production Number Purchase Date About Trouble in detail as possible Minimum Retention Period of Performance Parts for Repair The m...

Page 42: ...al functions Lock function Auto recovering after power failure Calibration offset Self diagnostic functions Failure of Sensor heater SSR memory internal communication temperature inputting circuit aut...

Page 43: ...e Symbol Part name ELB Earth leakage breaker CONT Control board TM Terminal block PIO Display circuit board H1 2 Heater TH1 Sensor for control X Main relay TH2 Sensor for overheating prevention SSR Br...

Page 44: ...ircuit board VS4 Yamato Scientific 1020000051 Tough card 50mm Yamato Scientific 1130000009 Main relay AHE1254 100V 120V Matsushita 2050000019 SSR TRS5225 Toho Denshi 2160000035 Power cord kit 2 0sq 3p...

Page 45: ...te Ammonium perchlorate and other perchlorate Potassium peroxide Sodium peroxide Barium peroxide and other inorganic peroxide Potassium nitrate Sodium nitrate Ammonium nitrate and other nitrate Sodium...

Page 46: ...43...

Page 47: ......

Page 48: ...document forbids Unexpected accidents or breakdown may result in Note The contents of this document may be changed in future without notice Any books with missing pages or disorderly binding may be re...

Reviews: