10.Specifications
Type
WL200
WL220
P
erfor
m
an
c
e
※
1
Sampling method
Ion exchange
Feed water type
One-touch coupler water supply connection resin hose
Sampled pure water
JIS K 0557
A3 deionized water
Sampling
rate
and
method
※
2
About 1L/min
deionized water, taken continuously
C
om
pos
iti
o
n
Ion
exchange
resin
cartridge
2L ion exchange resin
containing activated carbon
(CPC-P) 1 piece
2L ion exchange resin containing
activated carbon (CPC-P) 1 piece
2L ion exchange resin (CPC-E)
1
piece
Final filtration
0.1
μ
m membrane filter
Leak detection
Feed water electromagnetic valve forced to close on detecting leakage
S
tan
dard
Raw water pressure
range
0.5
~
5
×
100kPa
(
0.5
~
5kgf/
㎝
2
)
Safety devices
Circuit breaker, leak sensor, reducing valve, alarm of abnormal water
quality
Power
supply
(
50/60Hz
)※
3
Single phase
AC100
~
240V
0.05
~
0.2A
Overall dimensions
※
4
(
Width
×
Depth
×
Height
)
350mm
×
350mm
×
450
㎜
Sampling port
230mm above floor
Mass
About 14
㎏(
in dry state
)
About 16kg(in dry state)
Ma
rk
ing
Water quality display
LED display (conductivity/resistivity)
Other displays
Display for timing of consumables (CPC-P and CPC-E
:
to be replaced
simultaneously to prevent degradation of water quality)
Membrane filter, alarm display (water leakage alarm)
Accessories
Feed water hose set
Power cord
Instruction manual
Guaranty card
Ion exchange resin (CPC-P)
Ion exchange resin (CPC-E)
Seal tape
Membrane filter
1 piece
1 piece
This manual
1 piece
1 piece
1 piece (WL220 only)
1 piece
1 piece
Consumables
Name of part
Type
Article No.
Ion exchange resin cartridge (for
WL200)
CPC-P
253254
Ion exchange resin cartridge
(WL220
、
WL200 optional CPC-E
connection set (for OWL36)
※
5
CPC-P + CPC-E
253262
Membrane filter (two pieces
enclosed)
MFRL727
9020010004
※
1 The performance applies to the conditions with power supply of AC100
~
240V, room
temperature of 23
℃±
5
℃
, and humidity of 65
%
RH
±
20
%
.
The ambient temperature of this unit ranges from 5
℃
to 35
℃
. For the raw water temperature,
the guideline must be set to 10
℃~
35
℃
.
※
2 With the raw water of 1.0
×
100kPa (1kgf/cm
2
) or less, the sampling rate becomes 1.0L/min or
less.
※
3 The 100V power cord (with 3P plug attached) is provided.
※
4 No projection included
※
5 CPC-E and CPC-P are marketed as a set and not sold separately and individually.