24
Operation Method
Program Operation
4
Start operation
①
Press the ENTER key to decide the operation
start wait period/time. The unit enters to
program operation wait state. The blinking
PROGRAM lamp lights on and the STANDBY
lamp blinks instead in this state. The Sub
Display displays the step number and residual
time to operation start.
②
The operation starts when the residual time
counts zero. The STANDBY lamp lights off and
the Sub Display displays the executing step
number and the setting temperature after the
operation starts.
③
The following screens are displayed in sequence
during operation.
Set TEMP
37.0
℃↑
Step
S01
(Executing step number)
↓
Set TEMP
37.0
℃↑
Remain
15min
(Remaining time)
↓
Set TEMP
37.0
℃↑
Rep. count
10
(Residual count of repeat:
displayed during repeat only)
↓
Set TEMP
37.0
℃↑
Wait
5min
(Waiting state: displayed
during wait only)
④
The END lamp blinks and the Sub Display
displays the operation finish time when the
operation stops.
⑤
Press the POWER key to cancel the operation or
quit the wait state.