![Yamato IC103C Manual Download Page 11](http://html.mh-extra.com/html/yamato/ic103c/ic103c_manual_905240011.webp)
9
2. Before operating the unit
Installation procedures
・
precautions
(5) Take special care for samples shown below:
①
Samples that contain flammable or explosive components
・
The unit is not explosion proof. Never attempt to dry or process materials that contain
flammable or explosive components.
②
Corrosive samples
・
Take care for handling of corrosive samples. Although stainless steel is used for major
components, note that they might corrode with strong acid. Note that packing may
corrode with acid, alkali, oil or organic solvents.
③
Operation with devices with a larger heat load installed.
・
Note that temperature in the chamber may rise when operating the unit within a device.
(6) About exhaust ports.
・
IC103C/103CW /113C/113CW/403C/403CW/413C/413CW/603C/613C model
:
Located on
the top of the unit, IC803C/813C model
:
Located on either side of the unit.
Adjust the open amount according to the water content of a specific sample.
(IC103C/103CW/113C/113CW/403C/403CW/413C/413CW/603C/
613C model is shown in the diagram above.)
Adjust according to
water amount