32
In the Event of Failure
Safety Device and Error Code
This unit has an automatic diagnosis function built in the controller and safety devices independent of the
controller. The table below shows the cause and the solution method when the safety device operates.
Error Code:
When an abnormal condition occurs, an error code appears and the alarm lamp lights in the
controller, the buzzer sounds simultaneously. Record the error code and turn off the power of
device immediately.
Safety Device
Notify
Cause/Solution
Sensor trouble detection
“
ALARM
” lamp lights on,
“
Er.01
” appears
Temperature sensor is broken or
disconnected.
Make a call for service.
SSR short-circuit detection
“
ALARM
” lamp lights on,
“
Er.02
” appears
Triac is in short-circuit
Make a call for service.
Heater disconnecting detection
“
ALARM
” lamp lights on,
“
Er.03
” appears
Heater is disconnected.
Make a call for service.
Memory error
“
ALARM
” lamp lights on,
“
Er.15
” appears
Failure in internal memory.
Make a call for service.
Internal communication error
“
ALARM
” lamp lights on,
“
Er.17
” appears
Failure in internal communication or
temperature inputting circuit.
Make a call for service.
Overheating
“
ALARM
” lamp lights on,
“
Er.19
” appears
Overheating prevention device is in
operation.
Reset the power supply, and then
adjust the setting temperature of the
overheating protection device.
If the state does not recover, make a
call for service.
Measurement temperature
error
“
ALARM
” lamp lights on,
“
----
” appears
Measurement value is out of display
range.
Make a call for service.