14
DPI-500
Technical Manual
User Keyword Descriptions:
Appendix B: Error Codes
Keyword
A1
Baud Rate
A2
Data Bits and Parity
A3
Serial Transmission
Mode
A4
Display Check
A5
Disable the lb/kg
Key
A6
Serial Port Mode
A7
ID No. Enable
A8
ID No. Entry
A9
No. of Line
Feeds
Description
Select the baud rate for transmission through the serial port.
Selects number of data bits and parity for serial port transmission.
Selects the type of serial mode transmission.
"C" = Continuous mode; Send data continuously.
"d" = Demand mode; Send data on PRINT command from computer, printer, indicator.
Actuates the function which checks all display digits.
Allows user to disable the lb/kg key.
"0" = disable
"1" = enable
Selects the mode of the RS-232C serial port.
"0" = full duplex mode
"1" = Gross/Net/Tare print mode
Allows the ID No. to be disabled in the Gross/Net/Tare printout. Valid only when
A6
is set to "1".
"0" = disable
"1" = enable
Actuates the function which allows the entry of a new ID No. Valid only when
A6
is set to "1".
Pressing the
ZERO
key to scroll down once starts the sequence.
Actuates the function which allows entry of the desired number of line feeds to be printed after the
Gross/Net/Tare printout. Valid only when
A6
is set to "1". Pressing the
ZERO
key to scroll
down once starts the sequence.
Value
1200
2400
4800
9600
8n
7O
7E
C
d
Press ZERO key to
begin sequence
0
1
0
1
0
1
0
- 199999
0
- 99
Error
Err 0
Err 1
Err 2
Err 3
Err 4
Err 5
Err 7
Err 9
Where it Occurs
Normal Weighing Mode
Span Calibration Mode (F17)
Span Calibration Mode (F17)
Span Calibration Mode (F17)
All Modes
All Modes
Key-in Span Calibration Mode
(F20)
Initialization
Normal Weighing Mode
Meaning
Gross overload.
Keyed-in weight value in Calibration Procedure is less than 1% of full scale capacity.
Keyed-in weight value in Calibration Procedure is larger than full scale capacity.
Internal resolution is not high enough to process keyed-in weight value in Calibration
Procedure.
Diagnostics check error - EEPROM Read
Diagnostics check error - EEPROM Write
Attempted to key-in zero for C1.
No reading from analog to digital converter.
Span calibration lost.