YCO-1112-04-06
DP-6200
Technical Manual
The DP-6200 can be setup to be calibrated with pound weights or with kilogram weights. Check
the setting of system parameter 78 (see section IV - Changing the Scale Setup). 000 sets the scale
to be calibrated with kilogram weights, 001 sets the 60 lb and 300 lb scales to be calibrated with
pound weights, and 002 sets the 150 lb scale to be calibrated with pound weights.
1) Open the indicator housing by pulling the release tab forward and
swinging the front panel of the housing forward. Short the two test
pins to the right of the CPU on the main board. The Test Mode
ennunciator and internal counts will display.
2) Press and hold the
��
key and press the
���
�����
key. If the
scale indicates two point calibration mode, change it to three point
calibration mode by pressing the
���
�����
key.
If it becomes necessary to abort the calibration procedure before
completing it, press the
��
key to return to the test mode.
3) Remove all items from the scale platform, wait for the stable indi-
cator to display and press the
�
key.
4) Place 1/2 the full capacity of the scale on the platform, wait for the
stable indicator to display and press the
�
key.
5) Place the full capacity of the scale on the platform, wait for the
stable indicator to display and press the
�
key.
6) The scale will return to Test Mode and indicate full scale counts of
30000 for all kg, 27215 for 60 lb and 300 lb, and 34025 for 150 lb
calibration. The full scale counts should be /- 3 counts of
the appropriate target. Press the
���
key to exit Test Mode.
If the scale indicates error 103, then reconfirm that system pa-
rameter 78 is set correctly for the weights being used and the
capacity of the scale. Verify that the correct weights, 1/2 capac-
ity and full capacity, are being used. Make sure that nothing is
restricting the motion of the platform, pressing against the platform
or wedged between the upper and lower platform. Remove any
sources of vibration or electromagnetic radiation (i.e. - cell phones,
power cables, motors, compressors, etc.) Repeat the calibration
procedure. If the error repeats, see section V - Troubleshooting.
III Calibration
5
Test Mode
Two Point Mode
Three Point Mode
Zero Calibration
1/2 Capacity Calibration
Full Capacity Calibration
Full Scale Counts
Error 103