YCO-1112-04-06
DP-6200
Technical Manual
16
Key Function Description
Word
78 Conversion coefficient for 000 Kilogram calibration for all capacities.
pound calibration 001 Pound calibration for 60 lb and 300 lb capacity scales.
002 Pound calibration for 150 lb capacity scales.
79 Center-of-zero ennunciator 000 Alternate.
position toggle 001 NTEP.
80 Internal resolution 000 Do not use.
001~255
81 1 timer (ms) 000 Do not use.
001~255
82 3 timer (ms) 000 Do not use.
001~255
91 Mechanical zero 1 000~255 Automatically set at span adjustment.
92 Mechanical zero 2 000~255 Automatically set at span adjustment.
93 Span coefficient 1 (small) 000~255 Automatically set at span adjustment.
94 Span coefficient 2 (small) 000~255 Automatically set at span adjustment.
95 Span coefficient 1 (large) 000~255 Automatically set at span adjustment.
96 Span Coefficient 2 (large) 000~255 Automatically set at span adjustment.
97 Regional number and gravity 000~155 Automatically set at span adjustment.
for span adjustment
99 Factory setting 000 Always displayed. * Do Not Change *
IV Changing the Scale Setup
Parameter Descriptions - Factory Parameters (continued)