18
4. Operating procedures
Operating procedures (quick auto stop operation)
When you want to correct
set temperature or set time,
or change settings
When you want to change settings, press the
▼▲
keys on
the current screen to enter the setting mode where you can
change settings. Blinking stops in three seconds after
change and setting is completed. Note, however, that
temperature changes after timer activation are counted also
while temperature is changing.
When you want to change settings before timer activation,
press the TIMER
key on the current screen to enter the
setting mode where you can change settings. Enter a time
duration from when the set temperature is reached to the
time the device shall be stopped.
When you want to change settings after timer activation,
press the TIMER
key on the current screen to enter the
setting mode where you can change settings. Note,
however, you need to set a time calculated by adding the
time already passed to the time to be added.
After change has been made, press the RUN/STOP key to
complete the process.
When you want to stop quick auto stop operation in the
middle of it, press the RUN/STOP key long once to stop
device control once, then make settings again in the
appropriate mode.
In terms of the remaining time display
a blinking
dot indicates count down and an illuminating dot indicates a
wait status (while temperature is increasing or decreasing to
the set temperature) during which the timer has stopped
counting.