II Setting the Scale Functions
Technician Setup Mode
From the front, rotate the upper housing clockwise (to the right) until it rests on the table. Be careful not
to stress the wires that connect the main circuit board (in the upper housing) to the power board and
load cell (in the lower housing).
Place the platform back on the load cell assembly, power the scale on and press the tech ni cian setup
mode switch on the back of the main circuit board. The scale should indicate, “Set” followed by,
“Unit”.
The scale is now in the technician setup mode. From here, various settings can be altered and the
scale can be calibrated as detailed in “Setting Keywords” and in “Calibrating the Scale” below. Exit
the technician setup mode at any time by pressing the
ON/OFF
key for about fi ve seconds.
After exiting the technician setup mode, remove the platform from the load cell assembly, return the up-
per housing to the orig i nal position taking care not to pinch the wiring be tween the upper and lower
hous ing. Do not allow the wiring to rub against the upper load cell assembly. Replace the four up-
per housing screws and fi rmly tighten them. Place the scale back on it’s feet and place the platform
back on the scale.
Observe the supporting columns of the platform from the front and from the side. If the colums rub
against the upper housing, loosen the four upper housing screws, adjust the upper housing so that
it does not contact the platform columns and fi rmly tighten the four upper housing screws.
WARNING!
Do not place the scale on the platform or on the load cell assembly. This can over load
and damage the load cell.
2
Technician Setup
Mode Switch