20
Handling Precautions
About circulating liquid
・
Choose circulating liquid according to its operating temperature.
Target Temperature more than 10
℃:
City Water(or soften Industrial Water)
Target Temperature less than 10
℃:
Nonfreezing Solution(recommend Naiburain solution diluted by City
Water)
※
Never choose pure water and/or deionized water as circulating liquid.
Make the line up of concentration adjusted Naiburain solution so called Luck-rack Solution as
Yamato Products.
Please contact with local dealer or Yamato sales office and ask for Luck-rack Solution.
May cause failure of this Equipment/Unit so that pay attention to notices below.
・
May degrade the concentration percentage of nonfreezing solution for many months operation.
Check concentration percentage of nonfreezing solution and/or change the whole quantity of
nonfreezing solution by every 6(six) months as the standard.
・
Change the whole quantity of city water by every 3(three) months as the standard.
・
Never choose disqualified water(such as well water etc.), or pure water and deionized water as
circulating liquid.
・
May not meet with the Equipment/Unit performance because circulating pump will be
overworked, if apply high specific gravity and/or high viscosity liquid such as Fluorinert(by 3M)
and GALDEN(by Daitoku Tech) etc.
・
Never apply following solution to this Equipment/Unit.
1) Corrosive solution, and /or
2) Solution to generate corrosives at heating such as Fluorinert(by 3M) etc.
3) Solution to be harmed human body by absorbing its evaporating gas such as methanol
Check component materials to contact with solution below, choose adequate disinfectant, mix
it with circulating solution, and then apply.
(
Component materials to contact with solution:
Stainless Steel, Copper w/nickel plated, Silicon Rubber)
・
Be careful flammable solution to apply as nonfreezing solution such as ethanol, etc.
1) Ventilate laboratory/room well.
2) Never bring fire and igniting source(static electricity etc.) close to this Equipment/Unit.
Product Name
Product
Code
Concentration
(W t%)
STD Target
Temperature
Package Volume
( in litters)
Luck-rack
Solution: Z16005
756071
60
%
-20
℃
5L
Luck-rack
Solution: Z16010
756072
10L
Luck-rack
Solution: Z10005
756073
100
%
-30
℃
5L
Luck-rack
Solution: Z10010
756074
10L
Luck-rack
Solution: NFP6005
756075
60
%
-10
℃
5L
Luck-rack
Solution: NFP6010
756076
10L
About replacement of circulating liquid
Notice at changing from present circulating liquid to any other different kind of it. Must be drained
completely circulating liquid remained in its line of this Equipment and then change to new liquid.