4
Before Using this unit
Requirements for Installation
WARNING!
1. Always ground this unit
Connect the power plug to a receptacle with grounding connectors.
Do not forget to ground this unit, to protect you and the unit from electrical shock in case of
power surge. Choose a receptacle with grounding connectors as often as possible.
Do not connect the grounding wire to a gas pipe, or by means of a lightning rod or telephone
line. A fire or electrical shock will occur.
2. Choose a proper place for installation
Do not install this unit in a place where:
Rough or dirty surface.
Flammable gas or corrosive gas is generated.
Ambient temperature exceeds 35
C.
Ambient temperature fluctuates violently.
There is direct sunlight.
There is excessive humidity and dust.
There is a constant vibration.
Install this unit on a stable place with the space as shown below.
Main Unit
More than 15cm
Front side
3. Do not use this unit in an area where there is flammable or explosive gas
Never use this unit in an area where there
is flammable or explosive gas. This unit is
not explosion-proof. An arc may be
generated when the power switch is
turned ON or OFF, and fire/explosion may
result.
To know about flammable or explosive
gas refer to page20
”.
Summary of Contents for BS 200