Maintenance and electrode exchange
VARiON
®
Plus
700 IQ H
42
ba76207e02
12/2020
6.3
Exchanging the electrodes
Use the special socket wrench provided to dismantle an electrode.
Electrodes are installed as described in section 3.3.1 E
.
Recognizing the
electrode type from
outside
When mounted, the electrodes can be recognized by the following fea-
tures:
NOTICE
The sensor can be damaged by dirt and moisture. Each
time before dismantling an electrode carefully clean the
area around the electrodes (section 6.2). Before mount-
ing an electrode make sure the area behind the sealing
ring of the electrode and the receptacle are dry and
clean. The VARiON®Plus 700 IQ H may only be sub-
mersed when the electrodes or original blind plugs are
mounted.
Electrode
Hexagon
Front sur-
face
Other features
VARiON
®
Plus
Ref
black
black
– Thread at the
hexagon
VARiON
®
Plus
NH4
black
black
– Hexagon with-
out thread
VARiON
®
Plus
K
black
white
VARiON
®
Plus
NO3
white
white
– Labeled with the
electrode type
VARiON
®
Plus
Cl
black
black
– Larger mem-
brane
– Membrane flush
with the front
surface